Setup 12
EC Alarm
(Options /
Default
)
EC High Alarm
Enabled (En), Disabled (dIS)
(Enabled)
En
E C H I G H
dIS
E C H I G H
EC High Value
0.1 to 70.0 ppt *
(37.0 ppt)
0.1 to 70.0 PSU *
(37.0 PSU)
1.001 to 1.041 SG *
(1.028 SG)
37.0
E C H I G H
38.5
E C H I G H
With EC High Alarm configured as enabled “
EC HIGH VALUE
” is scrolled on the bottom of the LCD.
Actual EC value for high alarm (editable) is displayed blinking.
Press
e
to enter edit mode and use
a
/
v
to change value. Press
e
to save.
EC Low Alarm
Enabled (En), Disabled (dIS)
(Enabled)
En
E C L O W
dIS
E C L O W
EC Low Value
0.0 to 69.9 ppt *
(33.0 ppt)
0.0 to 69.9 PSU *
(33.0 PSU)
1.000 to 1.040 SG *
(1.025 SG)
33.0
E C L O W V
34.5
E C L O W V
With EC Low Alarm configured as enabled “
EC LOW VALUE
” is scrolled on the bottom of the LCD.
Actual EC value for low alarm (editable) is displayed blinking.
Press
e
to enter edit mode and use
a
/
v
to change value. Press
e
to save.
* Available range changes based upon other settings. High alarm value must be set higher than low alarm
(e.g. if low alarm was set 10.0 ppt, then high alarm range is 10.1ppt to 70 ppt).