
27
METHOD PROCEDURE
• Remove the cuvette.
• Add the content of one packet of
HI95771A-0
Total Chlorine UHR Reagent A and the content of
one packet of
HI95771B-0
Total Chlorine UHR
Reagent B. Replace the plastic stopper and the cap.
Shake gently for 20 seconds.
• Press
Read
. The display will show a 1 minute
countdown prior to the measurement. To skip the
timer, press
Read
twice. After 1 minute, invert the
cuvette 5 times.
• Insert the cuvette into the holder and ensure that
the notch on the cap is positioned securely in the
groove.
• Press
Read
twice. The instrument displays the results in
mg/L
of
chlorine (Cl
2
)
.
INTERFERENCES
Interference may be caused by:
• Bromine, Chlorine Dioxide, Chromium, Iodine, Oxidized Manganese, Ozone