METHOD PROCEDURE

• Add the content of one packet of 

HI93730A-0

 Molybdenum Reagent A to 

the cylinder, close and invert several times 
until completely dissolved.

• Add the content of one packet of 

HI93730B-0

 Molybdenum Reagent B to 

the cylinder, close and invert several times 
until completely dissolved.

• Add the content of one packet of 

HI93730C-0

 Molybdenum Reagent C to 

the cylinder, close and shake vigorously.

• Fill an empty cuvette with 10 mL of 

reacted sample (up to the mark). Replace 
the plastic stopper and the cap.

• Insert the cuvette into the holder and ensure 

that the notch on the cap is positioned securely 
in the groove.

10 mL

Содержание HI97730

Страница 1: ...INSTRUCTION MANUAL HI97730 Molybdenum Photometer...

Страница 2: ...a Hanna Instruments product Pleasereadthisinstructionmanualcarefullybeforeusingthisinstrument This manual will provide you with the necessary information for correct use of this instrument as well as...

Страница 3: ...METER VALIDATION CAL CHECK CALIBRATION 11 6 2 CHEMICAL FORMULA UNIT CONVERSION 12 6 3 GLP 13 6 4 LOGGING DATA LOG RECALL 13 6 5 GENERAL SETUP 14 6 6 REAGENTS ACCESSORIES 16 6 7 CONTEXTUAL HELP 16 6 8...

Страница 4: ...uvette A HI97730B CAL Check Cuvette B for Molybdenum Cloth for wiping cuvettes Scissors 1 5V AA Alkaline batteries CAL Check standard certificate Instrument quality certificate Instruction manual Each...

Страница 5: ...urs wipe up immediately and rinse with plenty of water If reagent contacts skin rinse the affected area thoroughly with water Avoid breathing released vapors Waste disposal For proper disposal of reag...

Страница 6: ...ight detector Silicon photocell Cuvette type Round 24 6 mm diameter 22 mm inside Additional Specifications Auto logging 50 readings Display 128 x 64 pixel B W LCD with backlight Auto off After 15 minu...

Страница 7: ...are able to validate the performance of the instrument at any time and apply a user calibration if necessary Hanna Instruments CAL Check cuvettes are made with NIST traceable standards The built in t...

Страница 8: ...ver Keypad Description The keypad contains 3 direct keys and 3 functional keys with the following functions Press the functional key to perform the function displayed above it on the LCD Press and hol...

Страница 9: ...sorbed by atoms molecules or crystal lattices Photometric chemical analysis is based on specific chemical reactions between a sample and reagent to produce a light absorbing compound If pure absorptio...

Страница 10: ...luminous efficiency providing more light while using less power They also produce little heat which could otherwise affect electronic stability LEDs are available in a wide array of wavelengths wherea...

Страница 11: ...m direct sunlight by keeping them in the original packing Store between 5 and 30 C 41 to 86 F do not freeze To perform a CAL Check 1 Press CAL Check from measurement mode The Not Available message or...

Страница 12: ...creen OUT OF SPECIFICATION A user calibration is not allowed the measured value is outside of the tolerance window Check the certified value expiration date and clean the outside of the cuvette Repeat...

Страница 13: ...l measurements Every time a measurement is made the data is automatically saved The data log can hold 50 individual measurements When the data log is full 50 data points the meter will rewrite the old...

Страница 14: ...ease or decrease the value Press Accept to confirm or the key to return to the Setup menu without saving the new value Contrast Option 0 to 100 Press Modify to change the display s contrast Use the fu...

Страница 15: ...ption Comma or Period Press the functional key to select the desired decimal separator The decimal separator is used on the measurement screen Language Press Modify to change the language Use the func...

Страница 16: ...ore Factory Settings Press Select to reset to factory settings Press Accept to confirm or Cancel to exit without restoring the factory settings 6 6 REAGENTS ACCESSORIES Press the key to enter the menu...

Страница 17: ...o diagnostic test when it is powered on During this test the Hanna Instruments logo will appear on the LCD If the auto diagnostic test was successful the meter is ready for use The battery icon on the...

Страница 18: ...tte a couple of times or for a specified time hold the cuvette in the vertical position Turn the cuvette upside down and wait for all of the solution to flow to the cap end then return the cuvette to...

Страница 19: ...s remove such bubbles by swirling or by gently tapping the cuvette Do not let the reacted sample stand too long after reagent has been added For best accuracy respect the timings described in the meth...

Страница 20: ...NT PROCEDURE Note If tutorial mode is disabled follow the measurement procedure below If the tutorial mode is enabled press Measure and follow the messages on the screen Fill the cuvette with 10 mL of...

Страница 21: ...Reagent B to the cylinder close and invert several times until completely dissolved Add the content of one packet of HI93730C 0 Molybdenum Reagent C to the cylinder close and shake vigorously Fill an...

Страница 22: ...ey and use the functional keys to select Chemical Form Use the functional keys and press Select to change the displayed chemical formula to mg L of molybdate MoO4 2 and sodium molybdate Na2MoO4 INTERF...

Страница 23: ...the zero cuvettes are inverted Swap the cuvettes and repeat the measurement There is either too much light or the instrument can not adjust the light level Please check the preparation of the zero cu...

Страница 24: ...e meter If the issue persists please contact Hanna Instruments technical support Battery level is too low for the meter to function properly Replace the batteries with new ones Tutorial mode has been...

Страница 25: ...e 11 ACCESSORIES 11 1 REAGENT SETS Code Description HI93730 01 Molybdenum Reagent 100 tests HI93730 03 Molybdenum Reagent 300 tests 11 2 OTHER ACCESSORIES Code Description HI7101412 blue carrying case...

Страница 26: ...hich it is used Any variation introduced by the user to the supplied equipment may degrade the meter s performance For yours and the meter s safety do not use or store the meter in hazardous environme...

Страница 27: ...Hanna Instruments reserves the right to modify the design construction or appearance of its products without advance notice...

Страница 28: ...World Headquarters Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com MAN97730 Printed in ROMANIA...

Отзывы: