21
METHOD PROCEDURE
• Remove the cuvette.
• Add 4 drops of
HI93715A-0
Ammonia MR
Reagent A. Replace the plastic stopper and the cap.
Invert 5 times to mix.
• Add 4 drops of
HI93715B-0
Ammonia MR
Reagent B. Replace the plastic stopper and the cap.
Invert 5 times to mix.
• Insert the cuvette into the holder and ensure that
the notch on the cap is positioned securely in the
groove.
• Press
Read
. The display will show a 3 minute and 30 second countdown prior to the
measurement. To skip the timer, press
Read
twice. When the timer ends, the meter
will perform the reading. The instrument displays the results in
mg/L
of
ammonia
nitrogen (NH
3
-N)
.
×
4
×
4