20
METHOD PROCEDURE
8. METHOD PROCEDURE
REQUIRED REAGENTS
Code Description
Quantity
HI93715A-0
Ammonia MR Reagent A
4 drops
HI93715B-0
Ammonia MR Reagent B
4 drops
REAGENT SETS
HI93715-01
Ammonia MR Reagent - 100 tests
HI93715-03
Ammonia MR Reagent - 300 tests
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
Note:
If tutorial mode is disabled, follow the measurement procedure below. If the
tutorial mode is enabled, press
Measure
and follow the messages on the screen.
• Fill the cuvette with 10 mL of unreacted sample
(up to the mark). Replace the plastic stopper and
the cap.
• Insert the cuvette into the holder and ensure that
the notch on the cap is positioned securely in the
groove.
• Press
Zero
. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
10 mL