31
Solution
for 15 minutes.
-
Inorganic
Soak in Hanna
HI7074 or HI8074 Inorganic
Cleaning Solution
for 15 minutes.
-
Oil/grease
Rinse with Hanna
HI7077 or HI8077 Oil and
Fat Cleaning Solution
.
IMPORTANT:
After performing any of the cleaning procedures rinse
the electrode thoroughly with distilled water, drain and refill the
reference chamber with fresh electrolyte, (not necessary for GEL filled
electrodes) and soak the electrode in
HI70300 or HI80300 Storage
Solution
for at least 1 hour before taking measurements.
TROUBLESHOOTING
Evaluate your electrode performance based on the following possibili-
ties.
• Noise
(Readings fluctuate up and down) could be due to:
- Clogged/Dirty Junction:
Refer to the Cleaning Procedure above.
- Loss of shielding
due to low electrolyte level (in refillable
electrodes only): refill with fresh
HI7071 or HI8071
for single
junction or
HI7082 or HI8082
for double junction electrodes.
• Dry Membrane/Junction:
Soak in
Storage Solution HI70300
or HI80300
for at least 1 hour.
• Drifting:
Soak the electrode tip in warm Hanna Solution
HI7082 or HI8082
for one hour then flush tip with distilled
water. Refill with fresh
HI7071 or HI8071
for single junction
electrodes and
HI7082 or HI8082
for double junction electrodes.
• Low Slope:
Refer to the cleaning procedure above.
• No Slope:
Check the electrode for cracks in glass stem or bulb
and replace the electrode.
• Slow Response/Excessive Drift:
Soak the tip in Hanna Solu-
tion
HI7061 or HI8061
for 30 minutes, rinse thoroughly in
distilled water and then follow the Cleaning Procedure above.