33
PU
m
P C
ALIB
rA
TION
Pr
OCEDU
rE
• Press
Start
, wait for the syringe refill.
• Place the stir bar in the beaker and put the beaker in the mini titrator top.
• Place the probe holder on the top of the beaker and secure it by turning clockwise.
• Rinse the pH electrode with deionized water and immerse into the sample until the reference
junction is completely submerged. Be sure that the tip of the electrode is not hitting the stir bar.
If necessary, additional distilled or deionized water can be added.
• Insert the dosing tip into the titrant tube sleeve. It is critical that the tip be immersed approximately
0.25 cm (0.1”) into the solution being titrated.
• Press
Continue
to begin the calibration and
Stop
to abort it.
• At the end of the calibration, “Calibration Completed“ appears on display. To repeat the calibration
press
Restart
or
ESC
to return to the main screen.