41
• Press
while in
ISE Measure
mode.
• Press
.
• Use
or
to select the
Calibration
option.
• Press
and use
or
to highlight the
Manual Entry
option.
• Press
or
to select the desired option.
Edit Custom Standards
Use
Edit Custom Standards
function to add additional
ISE standard values. Up to five custom standard values
can be added. Set Electrode Type and
Concentration Unit
prior to adding these standards.
To edit/set the
Custom Standards
:
• Press
while in
ISE Measure
mode.
• Press
.
• Use
or
to select the
Calibration
option.
• Press
and use
or
to
highlight the Edit Custom Standards option.
• If you want to disable the custom standard, press
. A pop-up menu will be displayed
asking for confirmation. Press
to confirm (the custom standard value will turn to “
----
”)
or press
to cancel the operation.
• Use
key to select the next custom standard to be set.
• Press
to return to
Edit Custom Standard
options.
ISE SETUP (HI5222 only)
Edit Standard Group
If a Group Standard was selected in the parameter Manual Entry, this parameter is used to create
your group of standards. If the Standard Group already contains five ISE standards, at least one
ISE standard has to be removed in order to add another standard.
To edit/set the
Standard Group
:
• Press
while in
ISE Measure
mode.
• Press
.
To set the
Manual Entry: