31
Log
Note: See section Logging for available types of logging.
This option allows the user to edit the logging settings: Logging Type, Logging Data Configuration,
Sampling Period and New Lot.
Logging Type
Three logging types are available: Automatic, Manual and Auto Hold.
Automatic
- the measurement data is logged automatically at predefined time intervals;
Manual
- a snapshot of the displayed measurement data is logged with time stamp when the user
manually presses Log;
Auto Hold
- this is configured along with the Direct/AutoHold Reading Mode to take a snapshot of
stable measurement data. Press
to initiate a logging session. Press
to initiate an
Auto Hold event. The log occurs automatically once measurement stability is reached. This type log
removes subjective data, as it only captures stable measurements.
To set the
Logging Type
:
• Press
while in
pH Measure
mode.
• Press
.
• Use
or
to select the
Log
option.
• Press
and use
or
to
highlight the
Logging Type
option.
• Press
and use
or
to
highlight the desired option.
• Press
to confirm your selection or press
to cancel operation.
Logging Data Configuration
This option allows the user to select which additional parameters will accompany a log report:
Date/Time, Calibration Data, Sample ID, Instrument ID, Operator ID, Company Name, Additional
Info 1 and Additional Info 2.
pH SETUP