9
Wiring & Controller Settings Configuration
3. WIRING & CONTROLLER SETTINGS CONFIGURATION
3.1. WIRING
Follow the lead markings (+ positive / – negative) to ensure that output leads are correctly wired to the
COMM
position on the main board.
COMM
- B A +
IN1 IN2 A01 A02 A03 A04
PROBE 2
- B A +
PROBE 1
- B A +
Terminal 1
Terminal 2
3.2. SETUP
• Press the
m
key to access the main menu.
• Press the
ud
keys to navigate to General setup.
• Enable
Remote Control
option.
• Press
Setup
to configure Communication Protocol parameters (to match the Modbus server).
Enable
b
Setup
3.2.1. Configurable parameters
Remote Control
Option: Enabled, Disabled
Option must be enabled when using the Modbus protocol.
The check mark confirms the option as enabled