Installation 8
Tank immersion with submersible electrode holder
I
II
III
1
2
4
3
5
6
7
9
8
The
HI60501
/
HI60503
(PVC / PVDF submersible holders) and
HI605011
(mounting flange) provide a sturdy
housing that prevents the probe from being damaged.
For high-temperature/complex chemical processes it is recommended the use of PVDF accessories.
To install the probe:
A. Unscrew the protective end cap (9).
The cap allows for quick and simple probe maintenance and replacement.
B. Slide out the electrode holder cap (3).
C. Screw the probe into the internally threaded protective cap (9), step
I
.
D. Screw the probe and protective cap subassembly onto the holder (5), step
II
.
E. The probe cable leads up through the holder (5) and out through the cap (3) and cable gland on top (2).
The cable is shielded inside the holder to prevent any damage to the insulation.
F. Tighten the cable gland (2), step
III,
and slide the cap (3) back onto the holder (5).
1 Probe cable
2 Cable gland
3 Electrode holder cap
4 Adjustable immersion level
5 Electrode holder
6 Probe body
7 O-ring, Ø 22.2 mm (0.87”)
8 Sensor tip
9 Protective end cap