Installation 8

Tank immersion with submersible electrode holder

 

I

II

III

1

2

4

3

5

6

7

9

8

The 

HI60501

 / 

HI60503

 (PVC / PVDF submersible holders) and 

HI605011

 (mounting flange) provide a sturdy 

housing that prevents the probe from being damaged. 
For high-temperature/complex chemical processes it is recommended the use of PVDF accessories.
To install the probe:

A.  Unscrew the protective end cap (9). 

The cap allows for quick and simple probe maintenance and replacement.

B.  Slide out the electrode holder cap (3).

C.  Screw the probe into the internally threaded protective cap (9), step 

I

.

D.  Screw the probe and protective cap subassembly onto the holder (5), step 

II

.

E.  The probe cable leads up through the holder (5) and out through the cap (3) and cable gland on top (2).

The cable is shielded inside the holder to prevent any damage to the insulation.

F.  Tighten the cable gland (2), step 

III,

 and slide the cap (3) back onto the holder (5).

1 Probe cable
2 Cable gland
3 Electrode holder cap
4 Adjustable immersion level
5 Electrode holder
6 Probe body
7 O-ring, Ø 22.2 mm (0.87”)
8 Sensor tip
9 Protective end cap

Содержание HI1006-18

Страница 1: ...rature Industrial Smart Probes SN XXXXXXXXXXXX pH 0 12 T 5 to 80 C p max 6 bar HI1006 1805 SN XXXXXXXXXXXX pH 0 12 T 5 to 80 C p max 6 bar HI1006 1800 Please scan the QR code or follow the link to dow...

Страница 2: ...ssories 13 13 Abbreviations 14 Certification 14 Recommendations for Users 14 Warranty 14 Dear Customer Thank you for choosing a Hanna Instruments product Please read this instruction manual carefully...

Страница 3: ...trial pH probes intended for industrial process control when teamed up with the HI510 Universal Process Controller HI1006 18 and HI1016 18 series designed for low conductivity or low temperature proce...

Страница 4: ...uoride resistant HF glass sensor titanium matching pin 5 0 to 60 0 C 23 0 to 140 0 F 0 00 to 10 00 pH 8 Smart probe with RS485 connection zz 00 supplied with DIN connector without cable See Accessorie...

Страница 5: ...controller setting Temperature compensation Automatic 0 0 to 80 0 C 32 0 to 176 0 F Temperature source Automatic from Probe Manual Body PVDF Junction PTFE Sensor Body LT glass Tip shape flat self cle...

Страница 6: ...ations Probes are easily installed using the 3 4 NPT external thread Hand tighten the probe in position Then depending on the process tighten one or two turns with a wrench to secure in place Do not e...

Страница 7: ...quire that saddle and flow cell are completely filled with water In line mounting with probe saddle 1 2 4 3 5 Note Probe can be rotated to prevent air entrapment inside the electrode Mounting accessor...

Страница 8: ...enance and replacement B Slide out the electrode holder cap 3 C Screw the probe into the internally threaded protective cap 9 step I D Screw the probe and protective cap subassembly onto the holder 5...

Страница 9: ...sediment or other foreign material to accumulate within the sensing area TANK VALVE 1 VALVE 2 FLOW FLOW INCOMING FLOW EXIT FLOW 3 1 2 4 5 7 6 6 FLOW CELL 1 Process controller 2 Wiring cable 3 pH prob...

Страница 10: ...tor and attach ground lead g COMM B A IN1 IN2 A01 A02 A03 A04 PROBE B A PROBE B A A B Probe cabling color code Marking Attached cable Patch cable Functionality GREEN BLACK 0V B WHITE WHITE RS485 D A Y...

Страница 11: ...CFM to save Wait followed by Calibration point is accepted and saved is displayed after the calibration point has been saved and accepted 5 Select Next to continue with a two point calibration 6 Repea...

Страница 12: ...ubbles inside the glass bulb If the bulb and or junction are dry soak the electrode in HI70300 Storage solution for at least 30 minutes To ensure a quick response time the glass bulb and the junction...

Страница 13: ...L Oil and fat cleaning solution 230 or 500 mL Other Accessories HI60501 PVC immersion electrode holder HI60503 PVDF immersion electrode holder HI60501 0 O rings for HI60501 electrode holder HI605011 P...

Страница 14: ...roduced by the user to the supplied equipment may degrade its performance Do not use or store the product in hazardous environments WARRANTY The probes are warrantied for six months against defects in...

Отзывы: