15. The LCD will display "Stor". The instrument checks the electrode

parameters and advises the user of abnormalities by "old probe"
and "dead probe" indications (in these cases, repeat the calibra-
tion with fresh buffers). If everything is satisfactory the meter is
calibrated and it returns to normal operational mode.

CALIBRATION ERROR MESSAGES

If the "old probe", "dead probe" or "WRONG" messages are displayed
during calibration, check your electrode by following the conditioning
and maintenance procedures and repeat calibration. The pH elec-
trode might have to be replaced if calibration cannot be successfully
performed.

note

: see "GLP" paragraph for more details about "old probe" and

"dead probe" messages.

Note

: The buffer pH value, and thus the value displayed on the

lower LCD, varies with temperature. For example, at 20

o

C it

shows  4.00 -7.03 -10.06,  at  25

o

C it shows 4.01-7.01-10.01.

6. Select the first buffer solution value with   and   if necessary.

7. When the "CFM" symbol blinks, the read-

ing is stable and calibration can be
confirmed.

8. Press ALT and CFM to confirm the first buffer.

9. If everything is satisfactory the LCD will

display "Stor" and then the second buffer
value expected will be displayed (two-
point calibration). If a wrong solution or
electrode has been used or if the buffer is
polluted, "WRONG" will be displayed to
alert the user.

10. If only a single point calibration is re-

quired, press CAL to exit the calibration
mode and maintain the previous slope
calibration. The instrument then checks
the electrode parameters and advises user
of abnormalities by "old probe" and "dead
probe" indications.

11. Press   or   to select the second buffer value. The meter will

display the "BUF 2" indication.

12. Rinse the electrode with some of the second buffer solution or clean

water.

13. Dip the bottom 4 cm (1½") of the pH electrode in a beaker

containing the second buffer.

14. When the "CFM" symbol blinks, press ALT

and CFM to confirm the second calibra-
tion point.

mV CALIBRATION PROCEDURE

(HI98150 ONLY)

A two or three-point calibration can be performed. The first two
calibration points are always 0 and +350 mV; the third point is
optional at +1900 mV.
1. Connect a mV simulator (

HI8427

 or 

HI931001

 with the proper

connecting cable) to the meter and set it to 0 mV.

2. Press CAL when LCD is displaying mV.
3. Enter the password (if different from "0000")

with the arrow keys.

4. Press ALT and CFM to confirm the password or CAL to exit.
5. If password is correct, the meter will

display "0 mV" on the lower LCD with the
"BUF 1" and "CAL" indication. The upper
LCD displays the mV reading.

6. When the "CFM" symbol blinks, the read-

ing is stable and calibration can be
confirmed.

7. Press ALT and CFM to confirm the first

value.

8. The LCD will display "Stor" for a few

seconds.

9. The meter will then display "350 mV" on

the lower LCD with the "BUF 2" indica-
tion.

Содержание HI 98140

Страница 1: ...nual HI 98140 HI 98150 Portable Waterproof pH ORP Meters with Smart Electrodes Log On Demand 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquip...

Страница 2: ...so incorporates a temperature sensor elimi nating the need for an additional temperature probe HI 98140 and HI 98150 feature all the most advanced features of laboratory pH meters Moreover they are pr...

Страница 3: ...ed data with ALT These instruments also feature a controlled access to calibration and GLP settings through a password protection method The Battery Error Preventing System BEPS detects when the batte...

Страница 4: ...re turning the meter off If the meter is OFF detach the electrode from the meter before immersing it in the storage solution ON OFF SPECIFICATIONS Range 4 00 to 19 99 pH 20 to 120 0 C HI 98150 only 60...

Страница 5: ...8 10 Calibration alarm time out 01 99 days OFF OFF 20 Auto Off Power down time out ON OFF ON 21 Firmware version 22 Battery level test 99 Password2 0000 9999 0000 Setup can be used to view data regard...

Страница 6: ...to change them pressing ALT and FNC with no need to enter setup mode Note If measurements are taken successively in different samples it is recommended to rinse the electrode thoroughly with deion iz...

Страница 7: ...om 0000 with the arrow keys 4 Press ALT and CFM to confirm the password or CAL to exit 5 If password is correct the meter displays 7 01pH on the lower LCD with the BUF 1 and CAL indication The upper L...

Страница 8: ...be displayed to alert the user 10 If only a single point calibration is re quired press CAL to exit the calibration mode and maintain the previous slope calibration The instrument then checks the ele...

Страница 9: ...tion at 0 0 50 0 C has to be performed in order to store the new calibration data in memory 1 Immerse the pH electrode with the built in temperature sensor in the 0 C temperature bath 2 Press ALT and...

Страница 10: ...hat an ORP electrode was connected mV reading HI98150 only means reading out of range temperature reading means reading out of range GLP message If RANGE is pressed when the GLP message is displayed t...

Страница 11: ...I98140 and HI98150 use electrodes with a built in EEPROM in which calibration data are read at start up and stored after calibration The meter can automatically analyze the data and advise the user if...

Страница 12: ...T The LCD can be illuminated to allow the user to see the readings even in dark environments It can be enabled disabled through the LIGHT key and it is automatically disabled when batter ies are in lo...

Страница 13: ...charged batteries The backlight can be enabled 2 Weakening batteries The backlight is automatically disabled and it is not possible to enable it until new batteries are inserted or an external power...

Страница 14: ...tuate up and down could be due to Clogged Dirty Junction refer to the cleaning procedure above Loss of shielding due to low electrolyte level for refillable electrodes only refill with fresh HI 7071 f...

Страница 15: ...ctrode Slow Response Excessive Drift soak the tip in HI 7061 solution for 30 minutes rinse thoroughly in distilled water and then follow the Cleaning Procedure above SMART ELECTRODES HI 1615D Smart pH...