• Press 

RANGE/GLP/

 to change the range.

The range can be changed any time when the
instrument is in 

measurement mode.

The selected range is memorized and the selection
remaines unchanged at power off or when the
battery is removed.

Code

Parameter

P1

Total Hardness

P2

Iron Low Range

The 

HI 96741

 can measure Free chlorine when range P1 is selected or Total chlorine when

range P2 is selected. To change the active range follow the procedure:

• Turn the meter on by pressing 

ON/OFF

. The

display briefly shows all tags on.

• After startup, the range identification number is

displayed on the secondary LCD as P1 or P2.

RANGE SELECTION

STARTUP

Prepare the instrument for measurement as follows:
• Unpack the instrument by removing the dust protection sleeve from the instrument cuvette

holder.

• Place the battery in the instrument as described in the “BATTERY REPLACEMENT” chapter.
• Place the instrument on a flat table.
• Do not place the instrument under direct sun light.

• It is important that the sample does not contain any debris.

This would corrupt the reading.

• Each time the cuvette is used, the cap must be tightened

to the same degree.

• Whenever the cuvette is placed into the measurement

cell, it must be dry outside, and completely free of
fingerprints, oil or dirt. Wipe it thoroughly with 

HI 731318

or a lint-free cloth prior to insertion.

• Shaking the cuvette can generate bubbles in the sample, causing higher readings. To obtain

accurate measurements, remove such bubbles by swirling or by gently tapping the cuvette.

• Do not let the reacted sample stand too long after reagent is added, or accuracy will be lost.

• It is possible to take multiple readings in a row, but it is recommended to take a new zero

reading for each sample and to use the same cuvette for zeroing and measurement.

• After the reading it is important to discard immediately the sample, otherwise the glass

might become permanently stained.

• All the reaction times reported in this manual are referred to 20°C (68°F). As a general rule

of thumb, they should be doubled at 10°C (50°F) and halved at 30°C (86°F).

• In order to maximize accuracy, prior to a measurement follow the 

validation procedure

 to

be sure that the instrument is properly calibrated. If necessary, calibrate the instrument.

Содержание HI 96741C

Страница 1: ...1 w w w h a n n a i n s t c o m Instruction Manual HI 96741C Total Hardness and Iron LR ISM...

Страница 2: ...LP 23 BATTERY MANAGEMENT 24 BATTERY REPLACEMENT 24 ACCESSORIES 25 WARRANTY 26 Dear Customer Thank you for choosing a Hanna product This manual will provide you with the necessary information for the c...

Страница 3: ...ive positive locking system to ensure that the cuvette is in the same position every time it is placed into the measurement cell It is designed to fit a cuvette with a larger neck making it easier to...

Страница 4: ...4 70 mg L Iron LR 0 00 to 1 60 mg L Resolution 0 01 mg L Total Hardness 0 01 mg L Iron LR Accuracy Mg Hardness 0 11 mg L 5 of reading 25 C Ca Hardness 0 11 mg L 5 of reading 25 C Iron LR 0 01 mg L 8...

Страница 5: ...oid measurement interference due to unwanted reflection and absorption of light It is recommended not to touch the cuvette walls with hands Furthermore in order to maintain the same conditions during...

Страница 6: ...t contain any debris Inverted cuvettes The sample and the zero cuvette are inverted Zero A zero reading was not taken Follow the instructions of the measurement procedure for zeroing the meter b on sa...

Страница 7: ...tes that the battery is dead and must be replaced Once this indication is displayed normal operation of the instrument will be interrupted Change the battery and restart the meter Under range A blinki...

Страница 8: ...s This would corrupt the reading Each time the cuvette is used the cap must be tightened to the same degree Whenever the cuvette is placed into the measurement cell it must be dry outside and complete...

Страница 9: ...seconds Fill a cuvette with 10 mL of the reacted sample up to the mark and replace the cap To compensate the meter for the sample turbidity or color the measurement takes place in two phases First th...

Страница 10: ...ve 4 0 mg L Nitrite ion above 0 8 mg L Chromium3 above 0 25 mg L Cobalt above 0 05 mg L Cyanide above 2 8 mg L Mercury above 0 4 mg L Nickel above 1 0 mg L Sample pH should be between 3 and 4 to avoid...

Страница 11: ...ate input CALIBRATION PROCEDURE Note It is possible to interrupt calibration procedure at any time by pressing CAL CHECK or ON OFF keys Warning Do not calibrate the instrument with standard solutions...

Страница 12: ...RY CALIBRATION RESTORE In GLP mode the last user calibration date can be verified and the factory calibration can be restored LAST CALIBRATION DATE GLP Press RANGE GLP to edit the desired year 2009 20...

Страница 13: ...ue is displayed when the instrument is switched on The blinking ZERO means that a new zero has to be performed One fresh battery lasts for around 750 measurements depending on the light level The rema...

Страница 14: ...intained according to the instructions This warranty is limited to repair or replacement free of charge Damages due to accident misuse tampering or lack of prescribed maintenance are not covered If se...

Страница 15: ...Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Technical Support for Customers Tel 800 426 6287 Fax 401 765 7575 E mailtech hannainst com www hannainst com...

Отзывы: