Hanna Instruments HI 96741C Скачать руководство пользователя страница 5

8

9

1) RANGE/GLP/  key
2) CAL CHECK key
3) ZERO/CFM key
4) READ/ /TIMER key

5) ON/OFF key
6) Liquid Crystal Display (LCD)

7) Cuvette alignment indicator

8) Cuvette holder

FUNCTIONAL DESCRIPTION

A microprocessor controlled special tungsten lamp emits radiation which is first optically
conditioned and beamed to the sample contained in the cuvette. The optical path is fixed by the
diameter of the cuvette. Then the light is spectrally filtered to a narrow spectral bandwidth, to
obtain a light beam of intensity 

I

o

 or 

I.

The photoelectric cell collects the radiation 

I

 that is not absorbed by the sample and converts

it into an electric current, producing a potential in the mV range.
The microprocessor uses this potential to convert the incoming value into the desired measuring
unit and to display it on the LCD.
The measurement process is carried out in two phases: first the meter is zeroed and then the
actual measurement is performed.
The cuvette has a very important role because it is an optical element and thus requires
particular attention. It is important that both, the measurement and the calibration (zeroing)
cuvettes, are optically identical to provide the same measurement conditions. Whenever possible
use the same cuvette for both. It is necessary that the surface of the cuvette is clean and not
scratched. This to avoid measurement interference due to unwanted reflection and absorption of
light. It is recommended not to touch the cuvette walls with hands.
Furthermore, in order to maintain the same conditions during the zeroing and the measuring
phases, it is necessary to close the cuvette to prevent any contamination.

INSTRUMENT DESCRIPTION

KEYPAD DESCRIPTION

ON/OFF

: to turn the meter on and off.

ZERO/CFM

: this is a bi-functional key. Just press to zero the meter prior to measurement,

or confirm edited values. In 

calibration mode press to confirm factory calibration restore.

READ/ /TIMER

: this is a multi-functional key. In 

measurement mode, press to make a

measurement, or press and hold for three seconds to start a pre-programmed countdown prior
to measurement. In 

GLP mode press to view the next screen.

CAL CHECK

: this is a bi-functional key. Just press to perform the validation of the meter, or

press and hold for three seconds to enter 

calibration mode.

RANGE/GLP/ : 

this is a multi-functional key. Just press to change the parameter. Press

and hold for three seconds to enter 

GLP mode. In calibration mode press to edit the date and time.

OPERATING MODES

Measurement mode: default operation mode, enables both 

validation 

and 

measurement

.

Calibration mode: may be entered by keeping 

CAL CHECK

 pressed 

for three seconds (

the

“CAL”

 tag appears), it enables 

calibration 

of the instrument.

GLP mode: may be entered by keeping 

RANGE/GLP/

 pressed for three seconds (

“GLP”

appears), it enables consulting of user calibration date or restore factory calibration.

Содержание HI 96741C

Страница 1: ...1 w w w h a n n a i n s t c o m Instruction Manual HI 96741C Total Hardness and Iron LR ISM...

Страница 2: ...LP 23 BATTERY MANAGEMENT 24 BATTERY REPLACEMENT 24 ACCESSORIES 25 WARRANTY 26 Dear Customer Thank you for choosing a Hanna product This manual will provide you with the necessary information for the c...

Страница 3: ...ive positive locking system to ensure that the cuvette is in the same position every time it is placed into the measurement cell It is designed to fit a cuvette with a larger neck making it easier to...

Страница 4: ...4 70 mg L Iron LR 0 00 to 1 60 mg L Resolution 0 01 mg L Total Hardness 0 01 mg L Iron LR Accuracy Mg Hardness 0 11 mg L 5 of reading 25 C Ca Hardness 0 11 mg L 5 of reading 25 C Iron LR 0 01 mg L 8...

Страница 5: ...oid measurement interference due to unwanted reflection and absorption of light It is recommended not to touch the cuvette walls with hands Furthermore in order to maintain the same conditions during...

Страница 6: ...t contain any debris Inverted cuvettes The sample and the zero cuvette are inverted Zero A zero reading was not taken Follow the instructions of the measurement procedure for zeroing the meter b on sa...

Страница 7: ...tes that the battery is dead and must be replaced Once this indication is displayed normal operation of the instrument will be interrupted Change the battery and restart the meter Under range A blinki...

Страница 8: ...s This would corrupt the reading Each time the cuvette is used the cap must be tightened to the same degree Whenever the cuvette is placed into the measurement cell it must be dry outside and complete...

Страница 9: ...seconds Fill a cuvette with 10 mL of the reacted sample up to the mark and replace the cap To compensate the meter for the sample turbidity or color the measurement takes place in two phases First th...

Страница 10: ...ve 4 0 mg L Nitrite ion above 0 8 mg L Chromium3 above 0 25 mg L Cobalt above 0 05 mg L Cyanide above 2 8 mg L Mercury above 0 4 mg L Nickel above 1 0 mg L Sample pH should be between 3 and 4 to avoid...

Страница 11: ...ate input CALIBRATION PROCEDURE Note It is possible to interrupt calibration procedure at any time by pressing CAL CHECK or ON OFF keys Warning Do not calibrate the instrument with standard solutions...

Страница 12: ...RY CALIBRATION RESTORE In GLP mode the last user calibration date can be verified and the factory calibration can be restored LAST CALIBRATION DATE GLP Press RANGE GLP to edit the desired year 2009 20...

Страница 13: ...ue is displayed when the instrument is switched on The blinking ZERO means that a new zero has to be performed One fresh battery lasts for around 750 measurements depending on the light level The rema...

Страница 14: ...intained according to the instructions This warranty is limited to repair or replacement free of charge Damages due to accident misuse tampering or lack of prescribed maintenance are not covered If se...

Страница 15: ...Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Technical Support for Customers Tel 800 426 6287 Fax 401 765 7575 E mailtech hannainst com www hannainst com...

Отзывы: