Please examine this product carefully. Make sure that the instrument is not damaged. If any
damage occured during shipment, please notify your Dealer.
Each 

HI 95762

 Portable Meter is supplied complete with:

• Two Sample Cuvets and Caps
• 9V Battery
• Scissors
• Two CAL CHECK™ Cuvets (HI 95762-11) with Quality Certificate
• Tissue for wiping cuvets
• Instruction Manual and Quick Reference Guide
• Instrument Quality Certificate
• Rigid carrying case

Note:

save all packing material until you are sure that the instrument works correctly. Any
defective item must be returned in its original packing.

PRELIMINARY EXAMINATION

18

REAGENT SETS
HI 95762-01

Reagents for 100 Free Chlorine tests

HI 95762-03

Reagents for 300 Free Chlorine tests

OTHER ACCESSORIES
HI 95762-11

CAL CHECK™ Standard Cuvets     (1 set)

HI 710009

Blue rubber boot

HI 710010

Orange rubber boot

HI 721310

9V battery (10 pcs)

HI 731318

Tissue for wiping cuvets (4 pcs)

HI 731331

Glass cuvets (4 pcs)

HI 731335

Caps for cuvets (4 pcs)

HI 93703-50

Cuvets cleaning solution (230 mL)

Battery replacement must only take place in a
non-hazardous area.
Simply slide off the battery cover on the back of
the meter. Detach the battery from the terminals
and attach a fresh 9V battery while paying attention
to the correct polarity. Insert the battery and replace
the cover.

9V

SLIDE

OFF

FACTORY CALIBRATION RESET

It is possible to restore factory calibration:
• Enter the 

calibration mode

 by holding CAL CHECK

for three seconds.

• Hold ZERO for 10 seconds. The display will

show for 2 seconds “F.CAL” and "762" appears.
The factory calibration is automatically restored
and the instrument is ready for measurement.

ACCESSORIES

BATTERY REPLACEMENT

Содержание HI 95762C

Страница 1: ...tion Manual HI 95762C Free Chlorine Meter for Drinking Water Application w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives 20 MAN95762R2 05 06 w w w h a n n a i n s...

Страница 2: ...ake all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid damages or burns do not put the ins...

Страница 3: ...ESSORIES HI 95762 11 CAL CHECK Standard Cuvets 1 set HI 710009 Blue rubber boot HI 710010 Orange rubber boot HI 721310 9V battery 10 pcs HI 731318 Tissue for wiping cuvets 4 pcs HI 731331 Glass cuvets...

Страница 4: ...62 portable meter measures the free chlorine Cl2 content in water samples in the 0 000 to 0 500 mg L ppm range The method is an adaption of the Standard Method 4500 Cl G for drinking water The reagent...

Страница 5: ...L chlorine and a representative lot of reagent an operator obtained with a single instrument a standard deviation of 0 004 mg L SPECIFICATIONS PRECISION AND ACCURACY 16 CALIBRATION PROCEDURE Turn the...

Страница 6: ...band interference filters to guarantee both high performance and reliable results HI 95 series block diagram optical layout PRINCIPLE OF OPERATION 15 Turn the meter on by pressing ON OFF When the LCD...

Страница 7: ...vet walls with hands Furthermore in order to maintain the same conditions during the zeroing and the measuring phases it is necessary to close the cuvet to prevent any contamination 1 Liquid Crystal D...

Страница 8: ...3 the Month and Day icons appear above the number of month and day when a date is displayed 4 the blinking Battery icon appears when the battery voltage is getting low 13 MEASUREMENT PROCEDURE Turn t...

Страница 9: ...ibration is selected GUIDE TO DISPLAY CODES 12 It is important that the sample does not contain any debris This would corrupt the reading Each time the cuvet is used the cap must be tightened to the s...

Страница 10: ...the right sample cuvet is inserted Inverted the sample and the zero cuvet are inverted Zero a zero reading was not taken Follow the instruction in the measurement procedure for zeroing the meter 11 U...

Отзывы: