background image

6

Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation and

matter. When a light beam crosses a substance, some of the radiation may be absorbed by atoms,
molecules or crystal lattices.
If pure absorption occurs, the fraction of light absorbed depends both on the optical path length through
the matter and on the physical-chemical characteristics of the substance according to the Lambert-
Beer Law:

-log

 

I

/

I

o

 = 

ε

λ

 

c d

or

A  

ε

λ

 

c d

Where:

-log

 

I

/

I

o

= Absorbance (A)

I

o

= intensity of incident light beam

I

= intensity of light beam after absorption

ε

λ

= molar extinction coefficient at wavelength

 

λ

c

= molar concentration of the substance

d

= optical path through the substance

Therefore, the concentration "c" can be calculated from the absorbance of the substance as the other
factors are known.
Photometric chemical analysis is based on the possibility to develop an absorbing compound from a
specific chemical reaction between sample and reagents. Given that the absorption of a compound
strictly depends on the wavelength of the incident light beam, a narrow spectral bandwidth should be
selected as well as a proper central wavelength to optimize measurements.
The optical system of Hanna's 

HI 95

 series colorimeters is based on special subminiature tungsten

lamps and narrow-band interference filters to guarantee both high performance and reliable results.

HI 95 series block diagram (optical layout)

PRINCIPLE OF OPERATION

15

• Turn the meter on by pressing ON/OFF.
• When the LCD displays “717”, it is ready.

• Place the CAL CHECK™ Standard Cuvet 

A

 into

the holder and ensure that the notch on the cap is
positioned securely into the groove.

• Press ZERO and “SIP” will blink on the display.

• After a few seconds the display will show “-0.0-”.

The meter is now zeroed and ready for validation.

• Remove the cuvet.

• Place the CAL CHECK™ Standard HI 95717-11

Cuvet 

B

 into the holder and ensure that the notch

on the cap is positioned securely into the groove.

• Press CAL CHECK and “SIP” will blink during

measurement.

Use the validation procedure to ensure that the instrument is properly calibrated.

Warning

: do not validate or calibrate the instrument with standard solutions other than the Hanna CAL

CHECK™ Standards, otherwise erroneous results will be obtained.

VALIDATION PROCEDURE

Temperature correction for Validation: corrected value = read value 

×

 [1 

+

 0.002 

×

 (20 

 T(°C))]

corrected value = read value 

×

 [1 

+

 0.001 

×

 (68 

 T(°F))]

• Wait for a few seconds and the display will show the validation standard value.

The reading should be within specifications as  reported on the CAL CHECK™ Standard Certificate.
If the value is found out of specifications, please check that the cuvets are free of fingerprints, oil or
dirt and repeat validation. If results are still found out of specifications then recalibrate the

instrument.

Содержание HI 95717C

Страница 1: ...1 Instruction Manual HI 95717C Phosphate High Range ISM w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives 20 MAN95717R2 08 06 w w w h a n n a i n s t c o m...

Страница 2: ...take all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid damages or burns do not put the i...

Страница 3: ...HI 95717 11 CAL CHECK Standard Cuvets 1 set HI 710009 Blue rubber boot HI 710010 Orange rubber boot HI 721310 9V battery 10 pcs HI 731318 Tissue for wiping cuvets 4 pcs HI 731331 Glass cuvets 4 pcs H...

Страница 4: ...he phosphate PO4 3 content in water samples in the 0 0 to 30 0 mg L ppm range The method is an adaption from Standard Methods for the Examination of Water and Wastewater The reagents are in powder and...

Страница 5: ...In a laboratory using a standard solution of 12 00 mg L phosphate and a representative lot of reagents an operator obtained with a single instrument a standard deviation of 0 5 mg L PRECISION AND ACC...

Страница 6: ...mps and narrow band interference filters to guarantee both high performance and reliable results HI 95 series block diagram optical layout PRINCIPLE OF OPERATION 15 Turn the meter on by pressing ON OF...

Страница 7: ...at the surface of the cuvet is clean and not scratched This to avoid measurement interference due to unwanted reflection and absorption of light It is recommended not to touch the cuvet walls with han...

Страница 8: ...display area 2 CAL appears when the instrument operates in the calibration mode 3 the Month and Day icons appear above the number of month and day when a date is displayed 4 the blinking Battery icon...

Страница 9: ...ation is selected GUIDE TO DISPLAY CODES 12 Proper use of the dropper a to get good reproducible results tap the dropper on the table for several times and wipe the outside of the dropper tip with a c...

Страница 10: ...0 indicates that the sample absorbs less light than the zero reference Check the procedure and make sure you use the same cuvet for reference zero and measurement A flashing value of the maximum conce...

Отзывы: