background image

3

PRELIMINARY EXAMINATION

Please examine this product carefully. Make sure that the instrument is not damaged. If any damage
occured during shipment, please notify your Dealer.
Each 

HI 95717

 Ion Selective Meter is supplied complete with:

• Two Sample Cuvets and Caps
• 9V Battery
• Scissors
• Two CAL CHECK™ Cuvets (HI 95717-11) with Quality Certificate
• Tissue for wiping cuvets
• Instruction Manual and Quick Reference Guide
• Instrument Quality Certificate
• Rigid carrying case

Note:

save all packing material until you are sure that the instrument works correctly. Any defective
item must be returned in its original packing.

18

REAGENT SETS
HI 93717-01

Reagents for 100 tests

HI 93717-03

Reagents for 300 tests

OTHER ACCESSORIES
HI 95717-11

CAL CHECK™ Standard Cuvets     (1 set)

HI 710009

Blue rubber boot

HI 710010

Orange rubber boot

HI 721310

9V battery (10 pcs)

HI 731318

Tissue for wiping cuvets (4 pcs)

HI 731331

Glass cuvets (4 pcs)

HI 731335

Caps for cuvets (4 pcs)

HI 93703-50

Cuvets cleaning solution (230 mL)

Battery replacement must only take place in a non-
hazardous area.
Simply slide off the battery cover on the back of the
meter. Detach the battery from the terminals and
attach a fresh 9V battery while paying attention to
the correct polarity. Insert the battery and replace
the cover.

9V

SLIDE

OFF

FACTORY CALIBRATION RESET

It is possible to restore factory calibration:
• Enter the 

calibration mode

 by holding CAL CHECK

for three seconds.

• Hold ZERO for 10 seconds. The display will show

for 2 seconds “F.CAL” and "717" appears. The
factory calibration is automatically restored and
the instrument is ready for measurement.

BATTERY REPLACEMENT

ACCESSORIES

Содержание HI 95717C

Страница 1: ...1 Instruction Manual HI 95717C Phosphate High Range ISM w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives 20 MAN95717R2 08 06 w w w h a n n a i n s t c o m...

Страница 2: ...take all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid damages or burns do not put the i...

Страница 3: ...HI 95717 11 CAL CHECK Standard Cuvets 1 set HI 710009 Blue rubber boot HI 710010 Orange rubber boot HI 721310 9V battery 10 pcs HI 731318 Tissue for wiping cuvets 4 pcs HI 731331 Glass cuvets 4 pcs H...

Страница 4: ...he phosphate PO4 3 content in water samples in the 0 0 to 30 0 mg L ppm range The method is an adaption from Standard Methods for the Examination of Water and Wastewater The reagents are in powder and...

Страница 5: ...In a laboratory using a standard solution of 12 00 mg L phosphate and a representative lot of reagents an operator obtained with a single instrument a standard deviation of 0 5 mg L PRECISION AND ACC...

Страница 6: ...mps and narrow band interference filters to guarantee both high performance and reliable results HI 95 series block diagram optical layout PRINCIPLE OF OPERATION 15 Turn the meter on by pressing ON OF...

Страница 7: ...at the surface of the cuvet is clean and not scratched This to avoid measurement interference due to unwanted reflection and absorption of light It is recommended not to touch the cuvet walls with han...

Страница 8: ...display area 2 CAL appears when the instrument operates in the calibration mode 3 the Month and Day icons appear above the number of month and day when a date is displayed 4 the blinking Battery icon...

Страница 9: ...ation is selected GUIDE TO DISPLAY CODES 12 Proper use of the dropper a to get good reproducible results tap the dropper on the table for several times and wipe the outside of the dropper tip with a c...

Страница 10: ...0 indicates that the sample absorbs less light than the zero reference Check the procedure and make sure you use the same cuvet for reference zero and measurement A flashing value of the maximum conce...

Отзывы: