45
Capacity: 8000 data samples which are di-
vided into 16 pages.
Each time a new logging period starts, it
automatically starts from a new page.
If the logging function is still on, and the
available page is 0 the meter will overwrite
the first lot of data in the existing memory.
During logging the meter automatically re-
turns to the oldest page in memory and if it
contains data, it will overwrite. In this case
the first log will not correspond to the oldest
set of data
It is recommended to periodically “clean” the
memory. Save data with PC if you need to
keep a record and then disconnect the bat-
teries for about 1 minute. If you do this,
remember to set the date and time, once the
batteries have been reconnected.
ATTENTION:
Data is stored into memory until the batteries
are removed.
If battery replacement is required and data is
not to be lost, plug the adapter in and pro-
ceed with battery replacement (see page 49).
Only once the batteries have been replaced it
is possible to unplug the adapter without
loosing the stored data.