
19
PROCEDURE FOR OFFSET CALIBRATION
(at 0.0 µS/cm)
• To perform the Offset
calibration, dry the con-
ductivity probe and leave
it in air.
• Press the ALT and the CAL keys simulta-
neously. The primary LCD display will blink
"
0.0
µS" for about 10 seconds.
• When "
0.0
µS" stops blinking, the cali-
bration can be confirmed. Press the CFM
key to confirm the offset value.
• If the input reading is out of the offset
range, an error message will show on the
secondary display "
E0
". Press the CAL
key to re-calibrate the value.
• If everything is satisfactory
the LCD will blink
"
84.0
µS", expecting the
first calibration solution.
• Press the CAL key to exit
the calibration mode now.
CAL
ALT
µ
S
µ
S
CFM
CAL
CAL
µ
S
Содержание HI 933300
Страница 56: ...h t t p w w w h a n n a i n s t c o m...