Power connection
Plug the 12VDC adapter (
HI710005
or
HI710006
) into the DC socket
(#7).
Plug the adapter into the mains.
Note:
The instrument uses the following configuration.
It is recommendable to use the Hanna
HI7710005
or
HI710006
voltage adapters (supplied witht the meters) that use the proper
polarity configuration.
The meters can also be used with other adapters. In this case, remem-
ber to check the correct polarity of your adapter before connecting it to
the meter.
Note:
Make sure the main line is protected by a fuse.
Note: HI9017, HI9218, HI9219, HI9318
use a rechargeable battery
to retain the pH calibration data only. It lasts approximately one
month when fully charged. The battery is automatically charged
any time the meter is plugged-in (either turned on or off).
Notes:
1)
If you need to unplug the instrument, press the OFF key before
disconnecting the meter from the mains. Only in this case will
the instrument retain the memorized data. Otherwise, the in-
strument needs to be re-calibrated. In the case of a power
failure while the instrument is operational , re-calibration is
required.
2)
When the instrument is initially switched on, the microprocessor
checks if any calibration data is in memory (i.e. if the offset is
within ±1 pH and the slope between 70 - 105%). If the
calibration data are not memorized, the offset and slope of the
instrument will be set to 0 mV and 100% respectively.