TO STOP LOGGING

Press the ALT and LOG keys simultaneously. This will also generate an
exit status printout.

TO SCAN STORED DATA ON DISPLAY

Press the LOG key. The log number and remaining page number will
appear on display.

While pressing the ALT key, press the DOWN key until the log number
to scan appears on the secondary display. The primary display will
show the number of samples in that particular log.

Press the ALT and RANGE keys simultaneously and the LCD will show
the logging starting date.

Press UP and the time of the most recent sample will be displayed.

Press UP and the temperature will be displayed on the primary LCD.

LOG

01-07-95

==02==     0005M

<>0010      *7.01

Log number

Total sample number

LOG EXIT
Date/Time

Press UP and the RH% value will be displayed on the primary LCD.

Continue pressing UP will display one by one all the memorized data of
the same log in the above sequence, i.e. time, temperature, RH% value.
Press DOWN key to revert back to sampling time and scan the samples.
To exit from the recall mode, press the LOG key.

Note:

This mode will not alter data already present in the memory.

PRINTING STORED DATA

Once a log number is selected, all or part of
that logged section can be printed by press-
ing the ALT and PRINT keys. The printout will
contain all logged samples in that section
beginning with the selected sample number
without altering the memory content.

Note:

 It is always possible to print only the

displayed sample by pressing the PRINT
key only.

For example if 10 samples are stored in the logging section, use the
DOWN key to display sample #5.

At this point, sample #5 can be printed pressing the PRINT key,
while samples #5, 6, 7, 8, 9 and 10 are printed if pressing ALT
and PRINT keys simultaneously.

To stop the printer at any time, press ALT and PAPER together.

Note:

 Before proceeding with printing, make sure there is enough

paper for the data to be printed. If the paper runs out, the
meter will not advise the operator and the printouts could be
lost. If this happens, stop the printer by pressing ALT and PAPER
keys simultaneously. Data will be kept in memory. Insert a new
paper roll (see "Printer  Maintenance" section for details),  and
repeat the above instructions starting from the last printed
sample number.

PRINT

ALT

PRINT

Содержание HI 91610

Страница 1: ...9161C ON OFF PAPER RANGE RH 5 0 95 0 C 20 0 60 0 INTV instrum nts e UP PRINT DOWN RANGE LOG MICROPROCESSOR PRINTING THERMOHYGROMETER HI 9161 HI 91610 Printing and Logging Thermo Hygrometers w w w h a...

Страница 2: ...CIFICATIONS 8 INITIAL PREPARATION 9 OPERATIONAL GUIDE 10 PRINTING RECORDING WITH HI 9161 17 PRINTING LOGGING WITH HI 91610 19 CALIBRATION 24 DATA TRANSFER TO PC HI 91610 only 27 SELF DIAGNOSTIC FUNCTI...

Страница 3: ...ecording printing interval relative humidity and temperature at user selectable intervals Both RH and temperature are measured with 0 1 resolution HI 91610 combines the high accuracy and portability o...

Страница 4: ...terval 6 TIME key to display current time and printing interval or to set date time printing interval with ALT 7 LOG key to start stop logging with ALT 8 LCD Liquid Crystal Display 9 PAPER key to adva...

Страница 5: ...se However when in recording logging mode the meter will continue to monitor RH and temperature To reactivate the display press RANGE INITIAL PREPARATION ON OFF RANGE RANGE SPECIFICATIONS HI 9161C HI...

Страница 6: ...ff until moisture has completely evaporated To accelerate the evaporation process expose the humidity sensor to a current of air For any problem in taking measurements please contact your dealer or th...

Страница 7: ...ng The printing interval on the secondary LCD will blink Any interval can be selected from among 1 2 5 10 15 30 60 120 and 180 minutes by using the UP and DOWN keys Once the desired interval is select...

Страница 8: ...red in the memory even when the meter or display are switched off Note Each time the batteries are replaced the settings will need to be re entered This can be avoided by applying an external voltage...

Страница 9: ...YY d Printing interval in minutes e Time HH MM f RH value g Temperature value When the meter is in recording mode LOG is displayed on the secondary LCD with the RH value on the primary LCD PRINT LOG V...

Страница 10: ...will print a complete set of data and the LOG symbol will appear on the secondary LCD Press the ALT and PAPER keys at the same time and the LOG symbol on display will start to blink If no key is press...

Страница 11: ...DOWN TIME LOG ALT RANGE LOG TIME To reactivate the display press the TIME key Notes Once in the logging mode the interval cannot be changed Exit the logging mode press ALT and LOG together and then s...

Страница 12: ...the recall mode press the LOG key Note This mode will not alter data already present in the memory PRINTING STORED DATA Once a log number is selected all or part of that logged section can be printed...

Страница 13: ...Na Cl 75 4 CALIBRATION KIT 12 3 6 9 4 h CALIBRATION KIT Li Cl 11 1 Na Cl 75 4 12 3 6 9 4 h All HANNA thermo hygrometers have been precalibrated at the factory HANNA instruments uses state of the art...

Страница 14: ...0 you need a 3 5 drive and a couple of minutes to follow the short instructions conveniently printed on the disk label TIME Windows is a registered Trademark of Microsoft Co Excel Copyright of Microso...

Страница 15: ...s for about 1 minute Remember to reset time and date once the batteries have been reinserted ATTENTION If batteries need to be replaced and data is not to be lost power the meter through a 12 Vdc adap...

Страница 16: ...er the cover preventing the pa per from advancing Replace the battery cover and tighten the screws 1 5V 1 5V 1 5V 1 5V SCREW POTS RESET CHANGING THE INK CARTRIDGE When printouts become faint it might...

Страница 17: ...handle HI 762BL Liquid probe with 1 m 3 3 cable black handle HI 762BL 10 Liquid probe with 10 m 33 cable black handle HI 762L Liquid probe with 1 m 3 3 cable white handle HI 762L 10 Liquid probe with...

Страница 18: ...the probe socket on the meter If the reading differs by more than 0 4 C 0 8 F from the test key value the unit is due for recalibration Choose the proper test key according to your application needs H...

Страница 19: ...42 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 0...

Отзывы: