background image

14

15

With the sensor facing down, screw the
membrane cap clockwise. Some electro-
lyte will overflow.

The platinum cathode sensor should always be
bright and untarnished. If it is tarnished or
stained, which could be due to contact with
certain gases or extended use with a loose or
damaged membrane, the cathode sensor
should be cleaned.

Use a clean, lint-free cardboard or cloth. Rub the
cathode very gently  side to side  4-5 times. This
will be enough to polish and remove any stains
without damaging the platinum tip. Then, rinse
the probe with deionized or distilled water.
Install a new membrane and fill it with fresh
electrolyte following the instructions above.
Recalibrate the instrument.

Important: For accurate and stable measure-
ments, it is important that the surface of the
membrane is in perfect condition. This perme-
able membrane isolates the sensor elements
from the environment but allows oxygen  to
enter. If any dirt is deposited on the membrane,
rinse it carefully with distilled or deionized
water. If any  imperfection is observed, or any
damage is evident (such as wrinkles, tears or
holes), the membrane should be replaced. Make
sure that the O-Ring is properly seated in the
membrane cap.

PROBE & MEMBRANE MAINTENANCE

The oxygen probe body is made of reinforced
polypropylene for maximum durability.
A built-in thermistor sensor measures the tem-
perature of the sample. It is recommended to
keep the protective cap on the probe when the
it is not in use.

To replace the membrane or refill it with
electrolyte, proceed as follows:

Remove the protective
cap by gently twisting and
pulling it off the probe
(fig.1).

Unscrew the membrane
by turning it counterclock-
wise (fig.2)

Wet the sensor by soak-
ing the bottom (2.5 cm /
1") of the probe in HI7041S
electrolyte solution for 5
minutes.

Rinse a new membrane
(HI 76407A) with some
electrolyte while shaking
it gently. Refill with fresh
electrolyte.

Gently tap the sides of
the membrane with your
finger to ensure that no
air bubbles remain
trapped. Do not directly
tap the bottom as this
may damage the mem-
brane.

Make sure that the rub-
ber O-ring is seated prop-
erly inside the membrane
cap.

TWIST

AND

PULL

fig. 1

UNSCREW

fig. 2

Содержание HI 9143

Страница 1: ...ruction Manual HI 9143 HI 9145 Portable D O Meters in Water Resistant Casing HI 9143 Microprocessor Auto Cal Dissolved Oxygen Meter CAL ON OFF RANGE FACTOR SAMPLE HOLD ppm C w w w h a n n a i n s t c...

Страница 2: ...mediately notify your dealer Each meter is supplied complete with HI76407 4 DOprobewith4 m 13 cable 2 spare membranes with O rings Protectivecap HI 7041S electrolyte solution 30 mL Batteries 4 x 1 5V...

Страница 3: ...et for 12 Vdc power adapter 19 LOW BAT low battery indicator 20 SAMPLE measuring mode indicator 21 F1 altitude factor indicator 22 F2 salinity factor indicator 23 or ppm oxygen measure unit 24 Seconda...

Страница 4: ...the cap clockwise Some electrolyte will over flow When not in use protect the membrane with the supplied cap SPECIFICATIONS Shipping cap black red THEN SCREW BACK ON THIRD TAP SECOND FILL FIRST INSER...

Страница 5: ...table measurements with the same recurring degree of accuracy With the probe properly polarized oxygen is continually consumed bypassingthroughthe sensitivediaphragmanddissolvingintheelec trolyte solu...

Страница 6: ...e ambient temperature and the temperatureofthesample thelongeritwilltake the probe to acclimatized itself to the sample TAKINGMEASUREMENTS Make sure the meter has been calibrated and the protective ca...

Страница 7: ...1 8 7 8 3 8 0 7 7 7 3 7 1 6 8 6 6 6 3 6 1 6 0 5 7 5 6 5 4 1800 m 11 8 11 2 10 6 10 1 9 6 9 2 8 8 8 3 8 0 7 7 7 4 7 1 6 8 6 6 6 3 6 1 5 9 5 8 5 5 5 4 5 2 F 32 0 35 6 39 2 42 8 46 4 50 0 53 6 57 2 60 8...

Страница 8: ...tilled or deionized water If any imperfection is observed or any damage is evident such as wrinkles tears or holes themembraneshouldbereplaced Make sure that the O Ring is properly seated in the membr...

Страница 9: ...on the top of the instrument and plug the power adapter into the socket Note The instrument uses the following con figuration Use the HANNA adapters such as HI 710005 orHI710006 withproperpolarityconf...

Страница 10: ...rding to the instruc tions Probes are warranted for six months Thiswarrantyislimitedtorepairorreplacement freeofcharge Damagesduetoaccidents misuse tampering or lack of prescribed maintenance are not...

Страница 11: ...3 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 27...

Отзывы: