background image

8

9

For an accurate calibration, it is recom-
mended to wait an additional 5-10 minutes to
ensure optimum conditioning of the probe.

Remove the protective cap.

Press CAL. The disply shows
"----" together with the "CAL"
symbol to indicate that the
instrument is in calibration
mode.

The instrument will automati-
cally standardize itself to the
actual saturation value. After
approx. 1 minute it will dis-
play "100%" with "SAMPLE"
to indicate that the calibra-
tion is complete.

HI 9143 only: press FACTOR and ensure
that F1 and F2 are set to the appropriate
altitude and salinity values.

Notes: • The instrument must be calibrated

whenever the probe, membrane or
electrolyte is changed.

• To exit the calibration

mode at any time,
press CAL.

• The display may be

switched from readings
in % saturation to  ppm
(mg/L) by simply
pressing the RANGE
key. Recalibration is
not needed.

CALIBRATION

PROBE POLARIZATION

The probe is under polarization with a fixed
voltage of approximately 800 mV. Probe polar-
ization is essential for stable measurements
with the same recurring degree of accuracy.
With the probe properly polarized, oxygen is
continually "consumed" by passing through the
sensitive diaphragm and dissolving in the elec-
trolyte solution contained inside the probe.
If this operation is interrupted, the electrolyte
solution continues to be enriched with oxygen
until it reaches an equilibrium with the surround-
ing solution. Measurements taken with a non-
polarized probe give an oxygen level that indi-
cates that of the test solution as well as any
oxygen present in the electrolyte solution. This
reading is obviously incorrect.
The HANNA HI 9143 and HI 9145 oxygen meter
automatically polarize the probe each time they
are switched on.

Calibration is simple and is recommended every
time the meter is switched on.

Make sure the probe is ready for measure-
ment (see page 7), i.e. the membrane is
filled with electrolyte, probe is connected to
the meter and properly polarized.

Switch the meter on.

"COND" appears on the dis-
play to inform the user that
the probe is in auto-condi-
tioning (automatic polariza-
tion) mode.

Once "COND" disappears, the probe is po-
larized and the instrument can be calibrated.

ON/OFF

CAL

CAL

SAMPLE

%

°C

CAL

RANGE

Содержание HI 9143

Страница 1: ...ruction Manual HI 9143 HI 9145 Portable D O Meters in Water Resistant Casing HI 9143 Microprocessor Auto Cal Dissolved Oxygen Meter CAL ON OFF RANGE FACTOR SAMPLE HOLD ppm C w w w h a n n a i n s t c...

Страница 2: ...mediately notify your dealer Each meter is supplied complete with HI76407 4 DOprobewith4 m 13 cable 2 spare membranes with O rings Protectivecap HI 7041S electrolyte solution 30 mL Batteries 4 x 1 5V...

Страница 3: ...et for 12 Vdc power adapter 19 LOW BAT low battery indicator 20 SAMPLE measuring mode indicator 21 F1 altitude factor indicator 22 F2 salinity factor indicator 23 or ppm oxygen measure unit 24 Seconda...

Страница 4: ...the cap clockwise Some electrolyte will over flow When not in use protect the membrane with the supplied cap SPECIFICATIONS Shipping cap black red THEN SCREW BACK ON THIRD TAP SECOND FILL FIRST INSER...

Страница 5: ...table measurements with the same recurring degree of accuracy With the probe properly polarized oxygen is continually consumed bypassingthroughthe sensitivediaphragmanddissolvingintheelec trolyte solu...

Страница 6: ...e ambient temperature and the temperatureofthesample thelongeritwilltake the probe to acclimatized itself to the sample TAKINGMEASUREMENTS Make sure the meter has been calibrated and the protective ca...

Страница 7: ...1 8 7 8 3 8 0 7 7 7 3 7 1 6 8 6 6 6 3 6 1 6 0 5 7 5 6 5 4 1800 m 11 8 11 2 10 6 10 1 9 6 9 2 8 8 8 3 8 0 7 7 7 4 7 1 6 8 6 6 6 3 6 1 5 9 5 8 5 5 5 4 5 2 F 32 0 35 6 39 2 42 8 46 4 50 0 53 6 57 2 60 8...

Страница 8: ...tilled or deionized water If any imperfection is observed or any damage is evident such as wrinkles tears or holes themembraneshouldbereplaced Make sure that the O Ring is properly seated in the membr...

Страница 9: ...on the top of the instrument and plug the power adapter into the socket Note The instrument uses the following con figuration Use the HANNA adapters such as HI 710005 orHI710006 withproperpolarityconf...

Страница 10: ...rding to the instruc tions Probes are warranted for six months Thiswarrantyislimitedtorepairorreplacement freeofcharge Damagesduetoaccidents misuse tampering or lack of prescribed maintenance are not...

Страница 11: ...3 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 27...

Отзывы: