METHODS OPTIMIZATION
vessel or other internal components of the cell may not be titrated with the rest of the sample.
It is equally important that no drops remain on the tip of the needle. ‘Hanging drops’ will end
up on the bottom of the septum. This will result in false low results for the determination.
Liquid samples with high viscosity like honey can be added via a syringe without needle
through the sample port following the steps outlined above.
In some cases liquid samples may require one of the additional preparatory steps listed in
the sections that follow. Specific sample preparation instructions are included with each
standard method.
9.2.5
Sample Preparation Techniques
While many samples can be introduced directly into the titration vessel (see section
6.5
Sample Addition), others require preparatory steps. It is critical that samples are not
contaminated with additional water or lose water during the preparation phase.
The steps required for the most common sample preparation techniques are outlined below.
For detailed application-specific instructions, consult the instructions included with applicable
standard methods.
The
HI 903
provides options for the automatic calculation of samples prepared normally,
using external extraction and external dissolution.
9.2.5.1
Internal Extractions
Internal extractions are carried out using the ‘normal’ sample type option within the ‘sample
parameters menu’. This type of sample preparation is suitable for solid samples which release
their water relatively quickly (during the pre-analysis stir time) and exhibit limited or no
solubility in Karl Fischer solvents. Internal extraction should be used preferentially over
external extraction techniques because the extracted water is titrated immediately, which
favors complete extraction by Le Chatlier’s principle.
An outline of the general procedure follows:
1. Add methanol or an appropriate solvent to the titration cell and pre-titrate to
dryness.
2. Adjust the pre-analysis stir time to be sufficiently long enough to complete the
extraction. Appropriate set times will be sample and solvent specific. Consult an
applicable standard method or experiment by increasing the pre-analysis stir time and
titrating samples until the resulting water content no longer increases.
3. Reduce the samples to as fine of a powder as possible to ensure that sample water
is extracted quickly.
4. Add the sample to the titration vessel using the back weighing method.
9.2.5.2
Dilutions
It is very difficult to accurately add very small amounts of sample to the titration vessel. In
order to produce accurate and reproducible results, samples having water content greater
than 50% should therefore be diluted with a dry solvent before being introduced into the
Содержание HI 903
Страница 1: ...1 QUICK START GUIDE HI 903 KARL FISCHER VOLUMETRIC TITRATOR Revision 1 0 ...
Страница 4: ...4 QUICK START GUIDE ...
Страница 7: ...7 QUICK START GUIDE TITRATOR CONNECTIONS Front View Rear View ...
Страница 18: ...18 QUICK START GUIDE QS 903 02 11 ...
Страница 19: ...1 INSTRUCTION MANUAL HI 903 KARL FISCHER VOLUMETRIC TITRATOR Revision 1 00 ...
Страница 22: ......
Страница 24: ...1 2 INTRODUCTION ...
Страница 26: ...2 2 SETUP ...
Страница 30: ...2 6 SETUP 2 3 2 Titrator Rear View 2 3 3 Titrator Left side View ...
Страница 40: ...2 16 SETUP ...
Страница 42: ...3 2 USER INTERFACE ...
Страница 50: ......
Страница 52: ...4 2 GENERAL OPTIONS ...
Страница 64: ......
Страница 92: ......
Страница 94: ...6 2 TITRA TITRA TITRA TITRA TITRATION TION TION TION TION ...
Страница 106: ...6 14 TITRA TITRA TITRA TITRA TITRATION TION TION TION TION ...
Страница 108: ...7 2 AUXILIARY FUNCTIONS AUXILIARY FUNCTIONS AUXILIARY FUNCTIONS AUXILIARY FUNCTIONS AUXILIARY FUNCTIONS ...
Страница 118: ...8 2 MAINTENANCE PERIPHERALS ...
Страница 147: ...A1 1 APPENDIX 1 Appendix 1 Contents A1 TECHNICAL SPECIFICATIONS A1 3 ...
Страница 148: ...A1 2 APPENDIX 1 ...
Страница 152: ......
Страница 155: ...A3 1 APPENDIX 3 Appendix 3 Accesories A3 TITRATOR COMPONENTS A3 3 ...
Страница 156: ......
Страница 159: ...1 GENERAL APPLICATIONS BROCHURE HI 903 KARL FISCHER VOLUMETRIC TITRATOR Revision 1 00 ...
Страница 160: ......
Страница 161: ......
Страница 163: ......
Страница 165: ......
Страница 167: ......
Страница 169: ......
Страница 171: ......
Страница 173: ......
Страница 174: ......
Страница 175: ......
Страница 177: ......
Страница 179: ......
Страница 181: ......
Страница 183: ......
Страница 185: ......
Страница 187: ......
Страница 189: ......
Страница 191: ...1 TITRATION THEORY TITRATION THEORY Principles HI 903 KARL FISCHER VOLUMETRIC TITRATOR Revision 1 0 ...
Страница 192: ...2 TITRATION THEORY ...
Страница 194: ...4 TITRATION THEORY ...
Страница 206: ...16 TITRATION THEORY A B Figure 8 Potential glass electrode mV C ...
Страница 220: ...30 TITRATION THEORY Titr THEORY 903 02 11 ...