• Pour at least 8 cm (3¼") of conductivity
solution into a plastic beaker.
• Immerse the probe into the conductivity
solution. The holes on the sleeve must be
completely submerged in the solution.
• Tap the probe repeatedly on the bottom of the beaker and stir it to
ensure that no air bubbles remain trapped inside the sleeve.
• If the temperature of the probe is close to that of the solution, the
display of
HI 8931
will stabilize quickly and provide a temperature
compensated conductivity measurement. Allow a few minutes if there
is a temperature difference of about 5°C (9°F) or more for the ATC
circuitry to compensate completely.
•
For HI8936
Adjust the SLOPE trimmer of the
HI 8936
transmitter to read on the
ammeter:
"10.4 mA" using
HI 7034
and
HI 8936 A/AL
"14.304 mA" using
HI 7030
and
HI 8936 B/BL
"15.304 mA" using
HI 7031
and
HI 8936 C/CL
"10.72 mA" using
HI 7033
and
HI 8936 D/DL