28
Material needed:
•
HI 7635
conductivity probe
•
HI 8931
conductivity controller
•
HI 8936
conductivity transmitter
• A 20 mA f.s. ammeter (for transmitters without LCD)
• A reference conductivity meter with automatic temperature compen-
sation accurately calibrated (e.g.
HI 8733
).
PROCEDURE
• Connect the
HI 7635
probe to the
HI 8936
transmitter (see page 21
for details).
• Connect the
HI 8936
transmitter to the
HI 8931
controller (see page
22 for details).
• Connect the
HI 8931
controller to the mains (see page 22 for de-
tails).
Before proceeding with the calibration, make
sure that the meter is in measurement mode
("MEASURE" LED is on) and not in set mode.
• Connect the ammeter to the
HI 8936
transmitter to monitor the sig-
nal current (see
below picture).
CALIBRATION PROCEDURE OF
CALIBRATION PROCEDURE OF
CALIBRATION PROCEDURE OF
CALIBRATION PROCEDURE OF
CALIBRATION PROCEDURE OF
HI 8931 & HI 8936 WITH HI 7635
HI 8931 & HI 8936 WITH HI 7635
HI 8931 & HI 8936 WITH HI 7635
HI 8931 & HI 8936 WITH HI 7635
HI 8931 & HI 8936 WITH HI 7635