![Hanna Instruments HI 8510 Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/hanna-instruments/hi-8510/hi-8510_instruction-manual_590924054.webp)
52
ELECTRODE CONDITIONING
AND MAINTENANCE
PREPARATION
Remove the protective cap.
DO NOT BE ALARMED IF ANY SALT DE-
POSITS ARE PRESENT.
This is normal with electrodes and they will
disappear when rinsed with water.
During transport tiny bubbles of air may have
formed inside the glass bulb. The electrode
Reference
Filling Hole
Reference
Filling Hole
Fill Hole
Screw
Sensitive
Wire
Reference
Wire
Reference
Junction
Glass
Bulb
Glass
Bulb
Reference
Junction
Reference
Junction
Reference
Junction
Platinum or
Gold tip
Platinum or
Gold tip
Reference
Wire
Reference
Wire
Sensitive
Wire
Glass Body
pH Electrode
Plastic Body
pH Electrode
Plastic Body
ORP Electrode
Glass Body
ORP Electrode
Reference
Wire