background image

25

24

• Press 

CAL

 to access the titration calibration menu where electrode and pump calibration may be accessed.

• Press 

HELP

 to view the contextual help.

• Press 

Continue 

or

 ESC 

to skip the message and enter titrator main screen.

• Press 

Purge 

to begin a purge cycle.

The purging stops automatically after 5 minutes.

To stop purging at any time and return to the main screen press 

ESC

 or 

Stop

.

During a purge, a countdown timer displays remaining time in 5 minute purge.

• Press 

Pause

 to interrupt the purge process.

• Press

 Pause 

or 

Stop

 (by pressing the corresponding function key in the purge screen)

• after the first drops of fresh titrant appear at the dosing tip

• in case an error condition is observed (empty titrant bottle, tubes or dosing tip

disconnected, pump error)

• if you want to resume at a later time

If 

Pause 

is pressed the next screen is displayed:

Press 

Resume

 to continue purging.

PURGE

Purging should be performed:

• whenever the tube of the peristaltic pump is replaced;

• whenever the titrant is changed or a new bottle of titrant is used;

• before starting a pump calibration;

• before starting a lot of titrations.

In order to start purging press the 

Purge

 key from the titrator main screen. The purging stops

automatically after 5 minutes.

To access the 

Purge

 key follow the next steps.

• From the instrument main screen (pH meter screen) press “Titrator” function key.

The instrument will display the next screen if tutorial is enabled and if any of the following

conditions exist:

- the meter hasn’t been calibrated in 8.30 pH buffer

- the pH calibration has expired

- a pump calibration hasn’t been performed or more than 3 days have passed since the last

 pump calibration

PUMP TUBE INSTALLATION

To mount the new peristaltic pump tube set follow next steps:

• Position one peristaltic pump fixing ring on its location.

• Stretch the tube over the peristaltic pump rolls.

• Attach the second pump fixing ring on its location.

• Attach the tube to the reagent bottle. Choose which titrant

is required for expected range. See table in Acidity

Type section.

Note

: Purge the peristaltic pump until drops of reagent appears on the dosing tip by pressing the

PURGE 

key from the titrator main screen.

To remove the tube of the peristaltic pump follow next steps:

Caution

: Purge line with water to remove titrant solution from tube.

• Detach the tubes system from the reagent bottle.

• Grasp one fixing ring of the peristaltic pump tube.

• Pull the tube until the fixing rings are taken out from their location.

• Remove the other side of the tube.

Содержание HI 84430

Страница 1: ...1 Instruction Manual HI 84430 TOTAL ACIDITY MINITITRATOR pH METER for Water Analysis www hannainst com...

Страница 2: ...mail us at tech hannainst com This instrument is in compliance with directives TABLE OF CONTENTS PRELIMINARY EXAMINATION 4 GENERAL DESCRIPTION 4 SPECIFICATIONS 6 PRINCIPLE OF OPERATION 8 FUNCTIONAL AN...

Страница 3: ...precipitates solids suspensions oils samples containing polyvalent cations aluminum iron manganese chromium may react slowly with the titrant causing a drifting end point To eliminate or diminish the...

Страница 4: ...le acidity high range mg L 400 4000 mg L as CaCO3 meq L 8 80 meq L as CaCO3 Resolution Titratable acidity low range 0 1 mg L 0 1 meq L Titratable acidity high range 1mg L 1meq L Accuracy 5 of reading...

Страница 5: ...than 3 7 only total acidity can be determined The HI 84430 utilizes two different titrant solutions to cover the entire measurement range The titrantsusedhavedifferentconcentrations appropriateforeach...

Страница 6: ...titrator top Purge the titrant Calibrate the pump Prepare the sample Run a titration and log sample results 5 Functional key area 6 Indicates that the displayed value can be changed using ARROW keys 7...

Страница 7: ...ut saving the new value pH resolution Options 0 1 0 01 Press the displayed function key in order to change the pH resolution Options C F Press the function key in order to change the temperature unit...

Страница 8: ...es on the screen Press the function key to select this option Tutorial Option AM PM or 24 hours Press the functional key to select the new value Time format Press to highlight the value to be modified...

Страница 9: ...ial link between the instrument and a PC Both must use the same setting Baud rate Option 0 to 9999 This option is used to set the instrument s ID The instrument ID is used while exchanging data with a...

Страница 10: ...olution to be used for the first calibration point Put the first beaker with calibration buffer in the beaker holder Place the electrode holder on the top of the beaker and secure it by turning clockw...

Страница 11: ...heck if the correct buffer has been used or regenerate the pH electrodes by following the Cleaning Procedure see the pH Cleaning Procedure section If necessary change the buffer or the electrode When...

Страница 12: ...temperature reading is out of the defined temperature range of the buffer 0 to 45 C the Wrong Buffer Temperature error message will be displayed and the symbol C will blink on the display Calibration...

Страница 13: ...key from the titrator main screen The purging stops automatically after 5 minutes To access the Purge key follow the next steps From the instrument main screen pH meter screen press Titrator function...

Страница 14: ...he electrode holder on the top of the beaker and secure it by turning clockwise Immerse the pH electrode approximatively 2 cm 0 8 into the standard paying attention not to touch the stir bar Insert th...

Страница 15: ...the beaker and secure it by turning clockwise Insert the temperature probe into the dedicated guide tube Immerse the pH electrode approximatively 2 cm 0 8 into the sample to be tested paying attention...

Страница 16: ...t of range the symbol is displayed in front of the reading From the titrator main screen press Titration To enter titrator main screen from pH meter mode press Titrator and then Titration If the tutor...

Страница 17: ...a calibration hasn t been performed the message Not Calibrated will be displayed If GLP pump is pressed one of the next screens is displayed Press Delete to enter record deleting mode Press Delete Al...

Страница 18: ...ayed preceded by the symbol and the MTC message The manually set temperature can now be adjusted with the ARROW keys from 20 0 to 120 0 C pH MEASUREMENT The HI 84430 may be used as a pH meter for dire...

Страница 19: ...to return to the previous screen without deleting If the pH log is empty the message No Records will be displayed VIEW DELETE RECORDED pH DATA To view or delete previously logged pH records press MEN...

Страница 20: ...eference thermometer The unstable measurement symbol will be shown on the display until the reading becomes stable When the reading is stable and close to the selected calibration point the unstable m...

Страница 21: ...he HI 92000 Windows compatible software optional HI 92000 also offers graphing and on line help features Data can be exported to the most popular spreadsheet programs for further analysis To connect y...

Страница 22: ...arameters setting The answer string contains Instrument ID 4 chars Calibration alarm time out 2 chars Concentration unit 1 char 0 mg L 1 meq L Acidity type 1 char 0 Total LR 1 Total HR 2 Strong LR 3 S...

Страница 23: ...o s p u s e t a u t c u l f g n i d a e R e s i o n n w o d d n a n o i t c n u j y t r i d d e g g o l C l e v e l e t y l o r t c e l e w o L s e d o r t c e l e H p e l b a l l i f e r y l n o n o...

Страница 24: ...ake all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instrument EMC performance To avoid damages or burns do not put the inst...

Страница 25: ...l 210 823 5192 Fax 210 884 0210 Indonesia Tel 210 4584 2941 Fax 210 4584 2942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 98...

Отзывы: