9
Result screen
The ion concentration is displayed in g/L (ppt) K
+
. If the
concentration is out of range, the instrument displays the upper
range limit blinking.
Press
LOG
to save the measured concentration. A message
indicating the free log space will be displayed for a few seconds.
If the free log space is under 10% this message will be
permanently displayed.
Press
Start
to begin a new measurement cycle.
Wrong slope screen
This screen appears if the computed slope isn’t in the
acceptable range. The acceptable range is 30% to 140%
default slope (53mV/decade).
Press
Start
to begin a new measurement cycle.
Add standard screen
Add the corresponding standard. Press
READ
to start the next
step analysis if
“Reading confirm”
option is enabled in the
Setup
menu. If
“Reading confirm”
option is disabled,
the
instrument will start the next step analysis.
Wrong standard error screen
This screen is displayed if step 2 and 3 of the current reading is
out of ±2000mV range or if the difference between the stable
readings from the current and the previous step is too small
(±3mV) or, if in auto confirm mode, the standard addition is
not detected.
Wrong input error screen
This screen is displayed if the current reading is out of
±2000mV range.
Pause screen
Press
Resume
to continue the measurement process.
Press
Stir
to start/stop the stirrer.
Sample reading confirm screen
(only if this feature is enabled
in setup)
Press
CFM
to memorize the reading.