GUIDE TO DISPLAY CODES

After Timer is pressed, a blinking hourglass icon appears and the
display shows a 2 hours coundown. Also the Zero tag might blink
if no zero measurement has been made before. At the end of the
countdown an acoustic signal alerts the user that the timer is
finished.

Indicates that the instrument is in a ready state and waiting for the
next command (Timer or Zero).

These prompts indicate the type of power supply: “Line” (if the
external power supply is used) or the battery level.

This prompt appears for a few seconds each time the instrument is
turned ON.

The instrument is zeroed and a measurement can be made.

Indicates that the meter is making a measurement.

Indicates that the meter is performing a zero measurement. The
light intensity is automatically re-adjusted (auto-calibration features)
if necessary.

Batteries voltage is getting low and the batteries need to be
replaced.

20

TOTAL PHENOLS, WHITE WINE

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS
Range

0.000 to 0.750 g/L

Resolution

0.001

Precision

±0.015 @ 0.350 g/L

Light Source

Tungsten lamp with narrow band interference filter @ 610 nm

Method

The reaction between Phenols and the reagents causes a blue tint in the
sample.

REQUIRED REAGENTS

REQUIRED REAGENTS

REQUIRED REAGENTS

REQUIRED REAGENTS

REQUIRED REAGENTS

Code

Description

Quantity/test

HI 83742A-0

Acid Reagent

5 mL

HI 83742B-0

Folin & Ciocalteu’s Reagent

6 drops

HI 83742C-0

Carbonate Buffer

4.5 mL

For reagent set and other accessories see page 26.

SAMPLE PREPARATION

SAMPLE PREPARATION

SAMPLE PREPARATION

SAMPLE PREPARATION

SAMPLE PREPARATION

• Use the 5 mL syringe to add exactly 5 mL of

HI 83742A-0 reagent to an empty cuvet.

Note

: in order to measure exactly 5 mL of reagent

with the syringe, follow the instructions on page 11.

5 mL

• Use the 200 µL automatic pipette to add exactly 0,2

mL of white wine sample to an empty cuvet.
For a correct use of the automatic pipette please
follow the related Instruction Sheet.

Содержание HI 83742

Страница 1: ...Instruction Manual HI 83742 COLOR PHENOLS ISM for wine analysis w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives MAN83742R3 04 06 w w w h a n n a i n s t c o m...

Страница 2: ...tampering or lack of prescribed maintenance are not covered If service is required contact your dealer If under warranty report the model number date of purchase serial number and the nature of the f...

Страница 3: ...hazardous area The blinking will appear when the batteries power gets low When batteries are completely discharged 0 bAtt will appear and after two seconds the instrument is switched off Remove the b...

Страница 4: ...ecommends usingthespecialwinesolventtominimizepossibleerrorsduetodilution Phenoldeterminationofwine Phenoliccompoundsareimportantforseveralreasonssincethey i affectthecolorofthewine ii have anastringe...

Страница 5: ...ungstenicacid H3PW12O40 andphosphomolybdenicacid H3PMo12O40 to thequinoidform R O TheconcomitantreductionoftheFolin Ciocalteureagentcausesabluecolorin thesamplethatisproportionaltothetotalphenoliccont...

Страница 6: ...s colorimeters is based on special subminiature tungsten lamps and narrow band interference filters to guarantee both high performance and reliable results Block diagram optical layout 23 TOTAL PHENOL...

Страница 7: ...This to avoid measurement interference due to unwanted reflection and absorption of light It is recommended not to touch the cuvet walls with hands Furthermore in order to maintain the same conditions...

Страница 8: ...to 20V DC 2 5 Watt FUNCTIONAL DESCRIPTION FRONT REAR INSTRUMENT DESCRIPTION 21 Add 6 drops of HI 83742B 0 Folin Ciocalteu s Reagent to the cuvet Replace the cap and shake gently to mix x 6 Wait for 1...

Страница 9: ...replaced 20 TOTAL PHENOLS WHITE WINE SPECIFICATIONS SPECIFICATIONS SPECIFICATIONS SPECIFICATIONS SPECIFICATIONS Range 0 000 to 0 750 g L Resolution 0 001 Precision 0 015 0 350 g L Light Source Tungste...

Страница 10: ...e cuvet for reference zero and measurement Inverted the sample and the zero cuvet are inverted b on sample reading b on sample reading b on sample reading b on sample reading b on sample reading Indic...

Страница 11: ...hen keeping the syringe in vertical position above the cuvet push the plunger completely down into the syringe Now the exact amount of 5 mL has been added to the cuvet GENERAL TIPS FOR AN ACCURATE MEA...

Страница 12: ...nual are referred to 20 C 68 F As a general rule of thumb they should be doubled at 10 C 50 F and halved at 30 C 86 F 17 MEASUREMENT MEASUREMENT MEASUREMENT MEASUREMENT MEASUREMENT Press RANGE to sele...

Страница 13: ...nm Method Direct reading REQUIRED REAGENTS REQUIRED REAGENTS REQUIRED REAGENTS REQUIRED REAGENTS REQUIRED REAGENTS Code Description Quantity test HI 83742 0 Wine Solvent 1 8 mL HI 83742 3 Wine Solven...

Страница 14: ...the zero zero zero zero zero Insert the zero cuvet into the holder and close the lid Press ZERO and will blink on the display SAMPLE PREPARATION SAMPLE PREPARATION SAMPLE PREPARATION SAMPLE PREPARATIO...

Отзывы: