8

DISPLAY ELEMENTS DESCRIPTION

1) Four digit main display.
2) Battery icon: appears when the battery voltage is getting low.
3) The hourglass icon: appears during the countdown.
4) Status information.
5) Measurement unit.
6) Lamp status indicator.
7) Four digit secondary display.

1) Lid
2) Cuvet Holder
3) Liquid Crystal Display (LCD)

4) ON/OFF key, to turn the meter on and off
5) ZERO key, to zero the meter

6) RANGE key, to select parameter
7) READ/TIMER key, to perform measurement
8) Power Socket 12V to 20V DC 2.5 Watt

FUNCTIONAL DESCRIPTION

FRONT

REAR

INSTRUMENT DESCRIPTION

21

• Add 6 drops of HI 83742B-0 Folin & Ciocalteu’s

Reagent to the cuvet.

• Replace the cap and shake gently to mix.

x 6

• Wait for 1 minute and then use the 3 mL pipette to

add HI 83742C-0 Reagent C Carbonate Buffer, to
bring the volume up to the 10 mL mark.

• Turn the meter ON and press RANGE to select the

parameter code “P4” for total phenols, white wine
(see page 13). Then hold down READ/TIMER for
several seconds. The instrument will show the
countdown. Alternatively, wait for 2 hours.
The instrument gives an acoustic signal to alert the
user that the countdown is finished.

• This is the reacted sample.

#2

(Sample)

10 mL

Содержание HI 83742

Страница 1: ...Instruction Manual HI 83742 COLOR PHENOLS ISM for wine analysis w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives MAN83742R3 04 06 w w w h a n n a i n s t c o m...

Страница 2: ...tampering or lack of prescribed maintenance are not covered If service is required contact your dealer If under warranty report the model number date of purchase serial number and the nature of the f...

Страница 3: ...hazardous area The blinking will appear when the batteries power gets low When batteries are completely discharged 0 bAtt will appear and after two seconds the instrument is switched off Remove the b...

Страница 4: ...ecommends usingthespecialwinesolventtominimizepossibleerrorsduetodilution Phenoldeterminationofwine Phenoliccompoundsareimportantforseveralreasonssincethey i affectthecolorofthewine ii have anastringe...

Страница 5: ...ungstenicacid H3PW12O40 andphosphomolybdenicacid H3PMo12O40 to thequinoidform R O TheconcomitantreductionoftheFolin Ciocalteureagentcausesabluecolorin thesamplethatisproportionaltothetotalphenoliccont...

Страница 6: ...s colorimeters is based on special subminiature tungsten lamps and narrow band interference filters to guarantee both high performance and reliable results Block diagram optical layout 23 TOTAL PHENOL...

Страница 7: ...This to avoid measurement interference due to unwanted reflection and absorption of light It is recommended not to touch the cuvet walls with hands Furthermore in order to maintain the same conditions...

Страница 8: ...to 20V DC 2 5 Watt FUNCTIONAL DESCRIPTION FRONT REAR INSTRUMENT DESCRIPTION 21 Add 6 drops of HI 83742B 0 Folin Ciocalteu s Reagent to the cuvet Replace the cap and shake gently to mix x 6 Wait for 1...

Страница 9: ...replaced 20 TOTAL PHENOLS WHITE WINE SPECIFICATIONS SPECIFICATIONS SPECIFICATIONS SPECIFICATIONS SPECIFICATIONS Range 0 000 to 0 750 g L Resolution 0 001 Precision 0 015 0 350 g L Light Source Tungste...

Страница 10: ...e cuvet for reference zero and measurement Inverted the sample and the zero cuvet are inverted b on sample reading b on sample reading b on sample reading b on sample reading b on sample reading Indic...

Страница 11: ...hen keeping the syringe in vertical position above the cuvet push the plunger completely down into the syringe Now the exact amount of 5 mL has been added to the cuvet GENERAL TIPS FOR AN ACCURATE MEA...

Страница 12: ...nual are referred to 20 C 68 F As a general rule of thumb they should be doubled at 10 C 50 F and halved at 30 C 86 F 17 MEASUREMENT MEASUREMENT MEASUREMENT MEASUREMENT MEASUREMENT Press RANGE to sele...

Страница 13: ...nm Method Direct reading REQUIRED REAGENTS REQUIRED REAGENTS REQUIRED REAGENTS REQUIRED REAGENTS REQUIRED REAGENTS Code Description Quantity test HI 83742 0 Wine Solvent 1 8 mL HI 83742 3 Wine Solven...

Страница 14: ...the zero zero zero zero zero Insert the zero cuvet into the holder and close the lid Press ZERO and will blink on the display SAMPLE PREPARATION SAMPLE PREPARATION SAMPLE PREPARATION SAMPLE PREPARATIO...

Отзывы: