Instruction Manual
HI 83742
COLOR & PHENOLS
ISM
for wine analysis
w w w . h a n n a i n s t . c o m
This Instrument is in
Compliance with the CE Directives
MAN83742R3 04/06
Страница 1: ...Instruction Manual HI 83742 COLOR PHENOLS ISM for wine analysis w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives MAN83742R3 04 06 w w w h a n n a i n s t c o m...
Страница 2: ...tampering or lack of prescribed maintenance are not covered If service is required contact your dealer If under warranty report the model number date of purchase serial number and the nature of the f...
Страница 3: ...hazardous area The blinking will appear when the batteries power gets low When batteries are completely discharged 0 bAtt will appear and after two seconds the instrument is switched off Remove the b...
Страница 4: ...ecommends usingthespecialwinesolventtominimizepossibleerrorsduetodilution Phenoldeterminationofwine Phenoliccompoundsareimportantforseveralreasonssincethey i affectthecolorofthewine ii have anastringe...
Страница 5: ...ungstenicacid H3PW12O40 andphosphomolybdenicacid H3PMo12O40 to thequinoidform R O TheconcomitantreductionoftheFolin Ciocalteureagentcausesabluecolorin thesamplethatisproportionaltothetotalphenoliccont...
Страница 6: ...s colorimeters is based on special subminiature tungsten lamps and narrow band interference filters to guarantee both high performance and reliable results Block diagram optical layout 23 TOTAL PHENOL...
Страница 7: ...This to avoid measurement interference due to unwanted reflection and absorption of light It is recommended not to touch the cuvet walls with hands Furthermore in order to maintain the same conditions...
Страница 8: ...to 20V DC 2 5 Watt FUNCTIONAL DESCRIPTION FRONT REAR INSTRUMENT DESCRIPTION 21 Add 6 drops of HI 83742B 0 Folin Ciocalteu s Reagent to the cuvet Replace the cap and shake gently to mix x 6 Wait for 1...
Страница 9: ...replaced 20 TOTAL PHENOLS WHITE WINE SPECIFICATIONS SPECIFICATIONS SPECIFICATIONS SPECIFICATIONS SPECIFICATIONS Range 0 000 to 0 750 g L Resolution 0 001 Precision 0 015 0 350 g L Light Source Tungste...
Страница 10: ...e cuvet for reference zero and measurement Inverted the sample and the zero cuvet are inverted b on sample reading b on sample reading b on sample reading b on sample reading b on sample reading Indic...
Страница 11: ...hen keeping the syringe in vertical position above the cuvet push the plunger completely down into the syringe Now the exact amount of 5 mL has been added to the cuvet GENERAL TIPS FOR AN ACCURATE MEA...
Страница 12: ...nual are referred to 20 C 68 F As a general rule of thumb they should be doubled at 10 C 50 F and halved at 30 C 86 F 17 MEASUREMENT MEASUREMENT MEASUREMENT MEASUREMENT MEASUREMENT Press RANGE to sele...
Страница 13: ...nm Method Direct reading REQUIRED REAGENTS REQUIRED REAGENTS REQUIRED REAGENTS REQUIRED REAGENTS REQUIRED REAGENTS Code Description Quantity test HI 83742 0 Wine Solvent 1 8 mL HI 83742 3 Wine Solven...
Страница 14: ...the zero zero zero zero zero Insert the zero cuvet into the holder and close the lid Press ZERO and will blink on the display SAMPLE PREPARATION SAMPLE PREPARATION SAMPLE PREPARATION SAMPLE PREPARATIO...