A microprocessor controlled special tungsten lamp emits radiation which is first optically conditioned

and beamed to the sample contained in the cuvet. The optical path is fixed by the diameter of the

cuvet. Then the light is spectrally filtered to a narrow spectral bandwidth, to obtain a light beam of

intensity 

I

o

 or 

I.

The photoelectric cell collects the radiation 

I

 that is not absorbed by the sample and converts it

into an electric current, producing a potential in the mV range.

The microprocessor uses this potential to convert the incoming value into the desired measuring

unit and to display it on the LCD.

The measurement process is carried out in two phases: first the meter is zeroed and then the actual

measurement is performed.

The cuvet has a very important role because it is an optical element and thus requires particular

attention. It is important that both the measurement and the calibration (zeroing) cuvets are

optically identical to provide the same measurement conditions. Whenever possible use the same

cuvet for both. It is necessary that the surface of the cuvet is clean and not scratched. This to avoid

measurement interference due to unwanted reflection and absorption of light. It is recommended

not to touch the cuvet walls with hands.

Furthermore, in order to maintain the same conditions during the zeroing and the measuring

phases, it is necessary to close the cuvet to prevent any contamination.

US Environmental Protection Agency

degree Celsius

degree Fahrenheit

milligrams per liter. mg/L is equivalent to ppm (part per million)

milliliter

Liquid Crystal Display

EPA

:

°C

:

°F

:

mg/L

:

mL

:

LCD

:

ABBREVIATIONS

14

• Add 4 full spoons of reagent 

to each vial.

x4

#2

x4

#1

x4

#2

x4

#1

• Using the other supplied spoon, add 4 full spoons of

reagent

 D

 to each vial.

• Close the glass vials tightly with their caps and shake

both vials vigorously for 1 minute.

Note: block the cap with a finger during shaking!

#2

#1

Содержание HI 83740

Страница 1: ...Instruction Manual HI 83740 COPPER ISM for wine analysis w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives MAN83740R3 04 06 w w w h a n n a i n s t c o m...

Страница 2: ...you with the necessary information for the correct use of the instrument Please read it carefully before using the meter If you need additional technical information do not hesitate to e mail us at t...

Страница 3: ...date of purchase serial number and the nature of the failure If the repair is not covered by the warranty you will be notified of the charges incurred If the instrument is to be returned to Hanna Inst...

Страница 4: ...L copper becomes a metabolic toxin that inhibits or delays alcoholic fermentation and concentrations exceeding 1 mg L may be sensorial detected and should be avoided Other copper related problems can...

Страница 5: ...nitions In a laboratory using a standard solution of 0 50 mg L copper and a representative lot of reagent an operatorobtainedwithasingleinstrumentastandard deviation of 0 05 mg L PRECISION AND ACCURAC...

Страница 6: ...lts Block diagram optical layout 15 Press TIMER and the instrument will show the countdown or alternatively wait for 10 minutes leaving the vials capped and undisturbed During this period the color of...

Страница 7: ...important that both the measurement and the calibration zeroing cuvets are optically identical to provide the same measurement conditions Whenever possible use the same cuvet for both It is necessary...

Страница 8: ...n 7 READ key to perform measurement 8 Power Socket 12V to 20V DC 2 5 Watt FUNCTIONAL DESCRIPTION FRONT REAR INSTRUMENT DESCRIPTION 13 Turn the instrument on by pressing ON OFF MEASUREMENT PROCEDURE 20...

Страница 9: ...uting procedure of the wine sample in case the L Lo Low Light message appears use the pipette to fill the glass vials with 5 mL of wine sample both vials then fill the vials up to the 15 mL mark with...

Страница 10: ...lowed during testing to ensure best accuracy Use the supplied plastic pipette for adding the exact amount of wine sample 15 mL or reagents A or B up to the 20 mL mark to the graduated glass vials The...

Отзывы: