PRINCIPLE OF OPERATION

Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation and

matter. When a light beam crosses a substance, some of the radiation may be absorbed by atoms,

molecules or crystal lattices.

If pure absorption occurs, the fraction of light absorbed depends both on the optical path length

through the matter and on the 

physical

-chemical characteristics of the substance according to the

Lambert-Beer Law:

-log

 

I

/

I

o

 = 

ε

λ

 

c d

or

A  

ε

λ

 

c d

Where:

-log

 

I

/

I

o

= Absorbance (A)

I

o

= intensity of incident light beam

I

= intensity of light beam after absorption

ε

λ

= molar extinction coefficient at wavelength

 

λ

c

= molar concentration of the substance

d

= optical path through the substance

Therefore, the concentration "c" can be calculated from the absorbance of the substance as the

other factors are known.

Photometric chemical analysis is based on the possibility to develop an absorbing compound from

a specific chemical reaction between sample and reagents. Given that the absorption of a

compound strictly depends on the wavelength of the incident light beam, a narrow spectral

bandwidth should be selected as well as a proper central wavelength to optimize measurements.

The optical system of Hanna's 

HI 83000

 series colorimeters is based on special subminiature

tungsten lamps and narrow-band interference filters to guarantee both high performance and

reliable results.

Block diagram (optical layout)

15

• Press TIMER and the instrument will show the

countdown or, alternatively, wait for 10 minutes,

leaving the vials capped and undisturbed.

During this period the color of the upper layer (organic

phase) in vial #2 will turn purple if copper is

present.

After 10 minutes the instrument gives an acoustic

signal to alert the user that the countdown has finished.

• Remove the cap of vial #1. Use the 3 mL plastic

pipette to transfer the upper layer (organic phase) into

a cuvet. 

Ensure that at least 1/3 of the cuvet is

filled with organic solvent 

(see page 11).

If some wine is transferred too, this does not interferes

with the measurement.

Cap the cuvet. This is the 

zero 

(#1).

Note:

The upper layer is an emulsion of small wine drops

dispersed in the organic phase. Please note that an

emulsion is an unstable equilibrium that may separate

even after few minutes. It is therefore important to

measure both the zero and the sample immediately

after the countdown has finished.

In case the emulsion separates before measurements

can be made, we recommend to leave the vials standing

for at least 4 hours, allowing complete separation of

the emulsion and obtaining two clear solutions in the

cuvets. Since the developed color is very stable, the

cuvets may be left standing overnight to be read also

next morning.

• Remove the cap of vial #2. Use the 3 mL plastic

pipette to transfer the upper layer (organic phase) into

another empty cuvet (see page 11).

If some wine is transferred too, this does not interferes

with the measurement.

Cap the cuvet. This is the reacted 

sample 

(#2).

#2

(Sample)

#1

(Zero)

Содержание HI 83740

Страница 1: ...Instruction Manual HI 83740 COPPER ISM for wine analysis w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives MAN83740R3 04 06 w w w h a n n a i n s t c o m...

Страница 2: ...you with the necessary information for the correct use of the instrument Please read it carefully before using the meter If you need additional technical information do not hesitate to e mail us at t...

Страница 3: ...date of purchase serial number and the nature of the failure If the repair is not covered by the warranty you will be notified of the charges incurred If the instrument is to be returned to Hanna Inst...

Страница 4: ...L copper becomes a metabolic toxin that inhibits or delays alcoholic fermentation and concentrations exceeding 1 mg L may be sensorial detected and should be avoided Other copper related problems can...

Страница 5: ...nitions In a laboratory using a standard solution of 0 50 mg L copper and a representative lot of reagent an operatorobtainedwithasingleinstrumentastandard deviation of 0 05 mg L PRECISION AND ACCURAC...

Страница 6: ...lts Block diagram optical layout 15 Press TIMER and the instrument will show the countdown or alternatively wait for 10 minutes leaving the vials capped and undisturbed During this period the color of...

Страница 7: ...important that both the measurement and the calibration zeroing cuvets are optically identical to provide the same measurement conditions Whenever possible use the same cuvet for both It is necessary...

Страница 8: ...n 7 READ key to perform measurement 8 Power Socket 12V to 20V DC 2 5 Watt FUNCTIONAL DESCRIPTION FRONT REAR INSTRUMENT DESCRIPTION 13 Turn the instrument on by pressing ON OFF MEASUREMENT PROCEDURE 20...

Страница 9: ...uting procedure of the wine sample in case the L Lo Low Light message appears use the pipette to fill the glass vials with 5 mL of wine sample both vials then fill the vials up to the 15 mL mark with...

Страница 10: ...lowed during testing to ensure best accuracy Use the supplied plastic pipette for adding the exact amount of wine sample 15 mL or reagents A or B up to the 20 mL mark to the graduated glass vials The...

Отзывы: