USING LIQUID AND POWDER REAGENTS

• Proper use of the dropper:

(a) for reproducible results, tap the dropper on the table for several times and wipe  the outside of the

dropper tip with a cloth.

(b) always keep the dropper bottle in a vertical position while dosing the reagent.

• Proper use of the powder reagent packet:

(a) use scissors to open the powder packet;
(b) push the edges of the packet to form a spout;

(c) pour out the content of the packet.

USING CUVETTES

• Proper mixing of the cuvette is done by 

shaking the cuvette

, moving the cuvette up and down. The

movement may be gentle or vigorous. This mixing method is indicated with “shake gently” or “shake

vigorously”, and one of the following icons:

(a)

(b)

The instructions listed below should be carefully followed during testing to ensure most accurate results.
• Color or suspended matter in large amounts may cause interference, they should be removed by

treatment with active carbon and filtration.

• Ensure the cuvette is filled correctly: the liquid in the cuvette forms a convexity on the top; the bottom

of this convexity must be at the same level of the 10 mL mark.

COLLECTING AND MEASURING SAMPLES

• In order to measure exactly 0.5 mL of reagent with the 1 mL syringe:

(a) push the plunger completely into the syringe and insert the tip into the solution.
(b) pull the plunger up until the lower edge of the seal is exactly on the 0.0 mL mark.
(c) take out the syringe and clean the outside of the syringe tip. Be sure that no drops are hanging

on the tip of the syringe, if so eliminate them. Then, keeping the syringe in vertical position above

the cuvette, push the plunger down into the syringe until the lower edge of the seal is exactly on

the 0.5 mL mark. Now the exact amount of 0.5 mL has been added  to the cuvette, even if the

tip still contains some solution.

TIPS FOR AN ACCURATE MEASUREMENT

(c)

(a)

(b)

(c)

(a)

(b)

shake gently

shake vigorously

Содержание HI 83206

Страница 1: ...1 Instruction Manual HI 83206 Multiparameter Bench Photometer for Environmental Testing w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives...

Страница 2: ...tible Software helps users manage all their results Dear Customer Thank you for choosing a Hanna product Please read this instruction manual carefully before using the instrument This manual will prov...

Страница 3: ...are optically identical to provide the same measurement conditions Most of methods use the same cuvette for both so it is important that measurements take place in the same optical point The instrumen...

Страница 4: ...the select method menu Press to move up in a menu or a help screen to increment a set value to access second level functions Press to move down in a menu or a help screen to decrement a set value to a...

Страница 5: ...mounts may cause interference they should be removed by treatment with active carbon and filtration Ensure the cuvette is filled correctly the liquid in the cuvette forms a convexity on the top the bo...

Страница 6: ...The blank and the sample are prepared exactly in the same way only the blank is deionized water instead of sample A blank cuvette may be used more than once stability and storing conditions are descri...

Страница 7: ...g and deleting You can view and delete the data log by pressing the RCL key Deleting is based on the LIFO last in first out scheme Additionally you can delete the data records all at once CHEMICAL FOR...

Страница 8: ...enabled a short beep is heard every time a key is pressed A long beep alert sounds when the pressed key is not active or an error condition is detected Press the functional key to enable disable the...

Страница 9: ...eractive contextual help mode that assists the user at any time To access help screens pressHELP The instrument will display additional information related to the current screen To read all available...

Страница 10: ...ide various aliphatic and aromatic amines Ammonia LR SPECIFICATIONS Range 0 00 to 3 00 mg L Resolution 0 01 mg L Accuracy 0 04 mg L 4 of reading Typical EMC 0 01 mg L Deviation Light Source Tungsten l...

Страница 11: ...ette with 10 mL of unreacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Free Chlorine FREE CHLORINE 10 mL Ammonia LR Remove the cuvette Add 4 drops o...

Страница 12: ...ey The meter will show 0 0 when the meter is zeroed and ready for measurement Remove the cuvette Powder reagents procedure Add the content of one packet of HI 93701 0 DPD reagent Replace the cap and s...

Страница 13: ...instrument x 3 x 3 x 1 Total Chlorine TOTAL CHLORINE SPECIFICATIONS Range 0 00 to 3 50 mg L Resolution 0 01 mg L Accuracy 0 03 mg L 3 of reading Typical EMC 0 01 mg L Deviation Light Source Tungsten l...

Страница 14: ...will show 0 0 when the meter is zeroed and ready for measurement Remove the cuvette and add the content of one packet of HI 93723 0 reagent Replace the cap and shake vigorously for about 10 seconds 1...

Страница 15: ...the cap Place the cuvette into the holder and close the lid Press ZERO key The meter will show 0 0 when the meter is zeroed and ready for measurement Remove the cuvette and add the content of one pac...

Страница 16: ...ce the blank 1 into the holder and close the lid Press ZERO key The meter will show 0 0 when the meter is zeroed and ready for measurement Remove the blank 10 mL 1 1 Reinsert the cuvette into the inst...

Страница 17: ...te with 10 mL of unreacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Press ZERO key The meter will show 0 0 when the meter is zeroed and ready for m...

Страница 18: ...ion section see page 12 Fill the cuvette with 10 mL of unreacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Press ZERO key The meter will show 0 0 wh...

Страница 19: ...ed sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Press ZERO key The meter will show 0 0 when the meter is zeroed and ready for measurement Fill a gradua...

Страница 20: ...other accessories see page 73 MEASUREMENT PROCEDURE Select the Molybdenum method using the procedure described in the Method Selection section see page 12 Fill the cuvette with 10 mL of unreacted sam...

Страница 21: ...above 10 mg L Iron above 50 mg L Nickel above 50 mg L Nitrite as NO2 Sulfate above 200 mg L Highly buffered samples or with extreme pH may exceed the buffering capacity of the reagents Fill one gradua...

Страница 22: ...he reaction between nickel and the reagent causes a blue tint in the sample REQUIRED REAGENTS Code Description Quantity HI 93726 0 Powder reagent 1 packet REAGENT SETS HI 93726 01 Reagents for 100 tes...

Страница 23: ...ical EMC 0 001 mg L Deviation Light Source Tungsten lamp with narrow band interference filter 575 nm Method Adaptation of the PAN method The reaction between nickel and the reagents causes an orange t...

Страница 24: ...cedure described in the Method Selection section see page 12 Using the pipette fill the cuvette with 6 ml of sample up to half of its height and replace the cap Place the cuvette into the holder and c...

Страница 25: ...f one packet of HI 93708 0 reagent Replace the cap and shake gently until completely dissolved 10 mL Nitrate Replace the cap and immediately shake vigorously up and down for exactly 10 seconds Continu...

Страница 26: ...ark with 10 mL of unreacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Press ZERO key The display will show 0 0 when the meter is zeroed and ready fo...

Страница 27: ...93732B 0 Add more sample to fill the bottle completely Replace the cap again and ensure that a part of the sample spills over This is to make sure that no air bubbles have been trapped inside which co...

Страница 28: ...cuvette and add 5 drops of HI 93710 0 Phenol Red Indicator Replace the cap and mix the solution 10 mL x 5 Dissolved Oxygen Let the sample stand and the flocculent agent will start to settle After app...

Страница 29: ...3717B 0 Reagent B 1 packet REAGENT SETS HI 93717 01 Reagents for 100 tests HI 93717 03 Reagents for 300 tests For other accessories see page 73 MEASUREMENT PROCEDURE Select the Phosphate HR method usi...

Страница 30: ...Press ZERO key The display will show 0 0 when the meter is zeroed and ready for measurement Remove the cuvette and add the content of one packet of HI 93713 0 reagent Replace the cap and shake gently...

Страница 31: ...reacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Press ZERO key The display will show 0 0 when the meter is zeroed and ready for measurement Remove...

Страница 32: ...eagent Replace the cap and swirl the solution Wait for 4 minutes add the content of one packet of HI 93705B 0 Citric acid reagent and shake until it is completely dissolved Wait for 1 minute This is t...

Страница 33: ...e the cuvette and add the content of one packet of HI 93705C 0 Amino acid reagent and shake until it is completely dissolved Reinsert the cuvette into the instrument Press TIMER and the display will s...

Страница 34: ...uracy 0 005 mg L 10 of reading Typical EMC 0 001 mg L Deviation Light Source Tungsten lamp with narrow band interference filter 575 nm Method Adaptation of the PAN method The reaction between silver a...

Страница 35: ...reagent close the cylinder and invert several times to mix until completely dissolved Fill one cuvette with 10 mL of the reacted sample up to the mark 20 mL 10 mL Silver Press ZERO key The display wil...

Страница 36: ...be caused by Aluminum above 6 mg L Cadmium above 0 5 mg L Copper above 5 mg L Iron above 7 mg L Manganese above 5 mg L Nickel above 5 mg L Zinc Place the cap and insert the cuvette into the instrumen...

Страница 37: ...Zincon ACCESSORIES REAGENT SETS HI 93700 01 100 ammonia LR tests HI 93700 03 300 ammonia LR tests HI 93701 01 100 free chlorine tests powder HI 93701 03 300 free chlorine tests powder HI 93701 F 300...

Страница 38: ...r dealer or the Hanna Customer Service Center nearest to you To find the Hanna Office in your vicinity check our home page at www hannainst com RecommendationsforUsers Before using these products make...

Страница 39: ...567 e mail sales hanna co jp Korea Tel 02 2278 5147 Fax 02 2264 1729 e mail mccoyhan chollian net Malaysia Tel 603 5638 9940 Fax 603 5638 9829 e mail hannamal tm net my Norway Tel 23 3811 00 Fax 23 38...

Отзывы: