background image

2

3

PRELIMINARY EXAMINATION

Please examine this product carefully. Make sure that the instrument is not damaged. If any damage

occurred during shipment, please notify your Dealer.
Each Meter is supplied complete with:

• Three Sample Cuvettes and Caps
• Cloth for wiping cuvettes (4 pcs)
• Scissors
• AC/DC Power Adapter
• Instruction Manual

Note:

Save all packing material until you are sure that the instrument works correctly. Any defective item

must be returned in its original packing with the supplied accessories.

°C

:

degree Celsius

EPA

:

US Environmental Protection Agency

°F

:

degree Fahrenheit

g/L

:

grams per liter (ppt)

HR

:

high range

LR

:

low range

mg/L

:

milligrams per liter (ppm)

mL

:

milliliter

MR

:

medium range

μ

g/L

:

micrograms per liter (ppb)

PAN

:

1-(2-pyridylazo)-2-naphtol

TPTZ

:

2,4,6-tri-(2-pyridyl)-1,3,5-triazine

ABBREVIATIONS

GENERAL DESCRIPTION

HI 83206 

is a multiparameter bench photometer dedicated for Environmental Testing. It can measure 24

different methods using specific liquid or powder reagents. The amount of reagent is precisely dosed to

ensure maximum reproducibility.

HI 83206

 bench photometer can be connected to a PC via an USB cable. The optional 

HI 92000

Windows

®

 Compatible Software helps users manage all their results.

Dear Customer,
Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the

instrument. This manual will provide you with the necessary information for the correct use of the

instrument. If you need additional technical information, do not hesitate to e-mail us at [email protected].
This instrument is in compliance with   directives.

TABLE OF CONTENTS

PRELIMINARY EXAMINATION ............................................. 3
ABBREVIATIONS ............................................................... 3
GENERAL DESCRIPTION ..................................................... 3
SPECIFICATIONS .............................................................. 4
PRECISION AND ACCURACY ............................................... 4
PRINCIPLE OF OPERATION ................................................ 4
FUNCTIONAL DESCRIPTION ............................................... 6
NEED TO KNOW ............................................................... 7
TIPS FOR AN ACCURATE MEASUREMENT ........................... 8
HEALTH & SAFETY ..........................................................11
METHOD REFERENCE TABLE .............................................11
OPERATIONAL GUIDE .......................................................12
SETUP ........................................................................... 14
HELP MODE ................................................................... 16
AMMONIA MR ................................................................17
AMMONIA LR ................................................................. 19
FREE CHLORINE ............................................................. 21
TOTAL CHLORINE ............................................................24
CHROMIUM VI HR ..........................................................27
CHROMIUM VI LR ........................................................... 29
COLOR OF WATER ........................................................... 31
COPPER HR....................................................................33

COPPER LR .................................................................... 35
CYANURIC ACID ..............................................................37
MOLYBDENUM ............................................................... 39
NICKEL HR ..................................................................... 42
NICKEL LR ...................................................................... 44
NITRATE ......................................................................... 47
NITRITE HR .................................................................... 49
NITRITE LR ..................................................................... 51
DISSOLVED OXYGEN ........................................................53
pH ................................................................................ 55
PHOSPHATE HR ..............................................................57
PHOSPHATE LR............................................................... 59
PHOSPHORUS ............................................................... 61
SILICA ........................................................................... 63
SILVER .......................................................................... 66
ZINC.............................................................................. 69
ERRORS AND WARNINGS ................................................. 71
DATA MANAGEMENT ........................................................72
STANDARD METHODS ......................................................72
ACCESSORIES ................................................................73
WARRANTY .................................................................... 75
HANNA LITERATURE ........................................................75

All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna

Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.

Содержание HI 83206

Страница 1: ...1 Instruction Manual HI 83206 Multiparameter Bench Photometer for Environmental Testing w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives...

Страница 2: ...tible Software helps users manage all their results Dear Customer Thank you for choosing a Hanna product Please read this instruction manual carefully before using the instrument This manual will prov...

Страница 3: ...are optically identical to provide the same measurement conditions Most of methods use the same cuvette for both so it is important that measurements take place in the same optical point The instrumen...

Страница 4: ...the select method menu Press to move up in a menu or a help screen to increment a set value to access second level functions Press to move down in a menu or a help screen to decrement a set value to a...

Страница 5: ...mounts may cause interference they should be removed by treatment with active carbon and filtration Ensure the cuvette is filled correctly the liquid in the cuvette forms a convexity on the top the bo...

Страница 6: ...The blank and the sample are prepared exactly in the same way only the blank is deionized water instead of sample A blank cuvette may be used more than once stability and storing conditions are descri...

Страница 7: ...g and deleting You can view and delete the data log by pressing the RCL key Deleting is based on the LIFO last in first out scheme Additionally you can delete the data records all at once CHEMICAL FOR...

Страница 8: ...enabled a short beep is heard every time a key is pressed A long beep alert sounds when the pressed key is not active or an error condition is detected Press the functional key to enable disable the...

Страница 9: ...eractive contextual help mode that assists the user at any time To access help screens pressHELP The instrument will display additional information related to the current screen To read all available...

Страница 10: ...ide various aliphatic and aromatic amines Ammonia LR SPECIFICATIONS Range 0 00 to 3 00 mg L Resolution 0 01 mg L Accuracy 0 04 mg L 4 of reading Typical EMC 0 01 mg L Deviation Light Source Tungsten l...

Страница 11: ...ette with 10 mL of unreacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Free Chlorine FREE CHLORINE 10 mL Ammonia LR Remove the cuvette Add 4 drops o...

Страница 12: ...ey The meter will show 0 0 when the meter is zeroed and ready for measurement Remove the cuvette Powder reagents procedure Add the content of one packet of HI 93701 0 DPD reagent Replace the cap and s...

Страница 13: ...instrument x 3 x 3 x 1 Total Chlorine TOTAL CHLORINE SPECIFICATIONS Range 0 00 to 3 50 mg L Resolution 0 01 mg L Accuracy 0 03 mg L 3 of reading Typical EMC 0 01 mg L Deviation Light Source Tungsten l...

Страница 14: ...will show 0 0 when the meter is zeroed and ready for measurement Remove the cuvette and add the content of one packet of HI 93723 0 reagent Replace the cap and shake vigorously for about 10 seconds 1...

Страница 15: ...the cap Place the cuvette into the holder and close the lid Press ZERO key The meter will show 0 0 when the meter is zeroed and ready for measurement Remove the cuvette and add the content of one pac...

Страница 16: ...ce the blank 1 into the holder and close the lid Press ZERO key The meter will show 0 0 when the meter is zeroed and ready for measurement Remove the blank 10 mL 1 1 Reinsert the cuvette into the inst...

Страница 17: ...te with 10 mL of unreacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Press ZERO key The meter will show 0 0 when the meter is zeroed and ready for m...

Страница 18: ...ion section see page 12 Fill the cuvette with 10 mL of unreacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Press ZERO key The meter will show 0 0 wh...

Страница 19: ...ed sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Press ZERO key The meter will show 0 0 when the meter is zeroed and ready for measurement Fill a gradua...

Страница 20: ...other accessories see page 73 MEASUREMENT PROCEDURE Select the Molybdenum method using the procedure described in the Method Selection section see page 12 Fill the cuvette with 10 mL of unreacted sam...

Страница 21: ...above 10 mg L Iron above 50 mg L Nickel above 50 mg L Nitrite as NO2 Sulfate above 200 mg L Highly buffered samples or with extreme pH may exceed the buffering capacity of the reagents Fill one gradua...

Страница 22: ...he reaction between nickel and the reagent causes a blue tint in the sample REQUIRED REAGENTS Code Description Quantity HI 93726 0 Powder reagent 1 packet REAGENT SETS HI 93726 01 Reagents for 100 tes...

Страница 23: ...ical EMC 0 001 mg L Deviation Light Source Tungsten lamp with narrow band interference filter 575 nm Method Adaptation of the PAN method The reaction between nickel and the reagents causes an orange t...

Страница 24: ...cedure described in the Method Selection section see page 12 Using the pipette fill the cuvette with 6 ml of sample up to half of its height and replace the cap Place the cuvette into the holder and c...

Страница 25: ...f one packet of HI 93708 0 reagent Replace the cap and shake gently until completely dissolved 10 mL Nitrate Replace the cap and immediately shake vigorously up and down for exactly 10 seconds Continu...

Страница 26: ...ark with 10 mL of unreacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Press ZERO key The display will show 0 0 when the meter is zeroed and ready fo...

Страница 27: ...93732B 0 Add more sample to fill the bottle completely Replace the cap again and ensure that a part of the sample spills over This is to make sure that no air bubbles have been trapped inside which co...

Страница 28: ...cuvette and add 5 drops of HI 93710 0 Phenol Red Indicator Replace the cap and mix the solution 10 mL x 5 Dissolved Oxygen Let the sample stand and the flocculent agent will start to settle After app...

Страница 29: ...3717B 0 Reagent B 1 packet REAGENT SETS HI 93717 01 Reagents for 100 tests HI 93717 03 Reagents for 300 tests For other accessories see page 73 MEASUREMENT PROCEDURE Select the Phosphate HR method usi...

Страница 30: ...Press ZERO key The display will show 0 0 when the meter is zeroed and ready for measurement Remove the cuvette and add the content of one packet of HI 93713 0 reagent Replace the cap and shake gently...

Страница 31: ...reacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Press ZERO key The display will show 0 0 when the meter is zeroed and ready for measurement Remove...

Страница 32: ...eagent Replace the cap and swirl the solution Wait for 4 minutes add the content of one packet of HI 93705B 0 Citric acid reagent and shake until it is completely dissolved Wait for 1 minute This is t...

Страница 33: ...e the cuvette and add the content of one packet of HI 93705C 0 Amino acid reagent and shake until it is completely dissolved Reinsert the cuvette into the instrument Press TIMER and the display will s...

Страница 34: ...uracy 0 005 mg L 10 of reading Typical EMC 0 001 mg L Deviation Light Source Tungsten lamp with narrow band interference filter 575 nm Method Adaptation of the PAN method The reaction between silver a...

Страница 35: ...reagent close the cylinder and invert several times to mix until completely dissolved Fill one cuvette with 10 mL of the reacted sample up to the mark 20 mL 10 mL Silver Press ZERO key The display wil...

Страница 36: ...be caused by Aluminum above 6 mg L Cadmium above 0 5 mg L Copper above 5 mg L Iron above 7 mg L Manganese above 5 mg L Nickel above 5 mg L Zinc Place the cap and insert the cuvette into the instrumen...

Страница 37: ...Zincon ACCESSORIES REAGENT SETS HI 93700 01 100 ammonia LR tests HI 93700 03 300 ammonia LR tests HI 93701 01 100 free chlorine tests powder HI 93701 03 300 free chlorine tests powder HI 93701 F 300...

Страница 38: ...r dealer or the Hanna Customer Service Center nearest to you To find the Hanna Office in your vicinity check our home page at www hannainst com RecommendationsforUsers Before using these products make...

Страница 39: ...567 e mail sales hanna co jp Korea Tel 02 2278 5147 Fax 02 2264 1729 e mail mccoyhan chollian net Malaysia Tel 603 5638 9940 Fax 603 5638 9829 e mail hannamal tm net my Norway Tel 23 3811 00 Fax 23 38...

Отзывы: