Instruction Manual

HI 83141

Portable pH/mV/ºC 

meter

w w w . h a n n a i n s t . c o m

WARRANTY

All Hanna Instruments

 meters are warranted for two years 

against

defects in workmanship and materials when used for their intended pur-
pose and maintained according to instructions.

The electrodes and the probes are warranted for a period of six
months. 

This warranty is limited to repair or replacement free of charge.

Damages due to accident, misuse, tampering or lack of prescribed mainte-
nance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of purchase,
serial number and the nature of the failure.
If the repair is not covered by the warranty, you will be notified of the
charges incurred. If the instrument is to be returned to Hanna Instruments,
first obtain a Returned Goods Authorization Number from the Customer
Service department and then send it with shipment costs prepaid.
When shipping any instrument, make sure it is properly packaged for
complete protection.

Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using the
meter. If you need additional technical information, do not
hesitate to e-mail us at

 [email protected]

.

This instrument is in compliance with the 

 directives.

Remove the instrument from the packing material and exam-
ine it carefully to make sure that no damage has occurred
during shipment. If there is any damage, notify your Dealer.
The meter is supplied complete with

• HI 1230B

 plastic-body, double-junction, gel, com-

bination pH electrode with BNC and 1 m cable

• HI 7669AW

 temperature probe with 1 m cable

• 1 x 9V battery, screwdriver and instructions.

Note:

Conserve all packing material until the instrument has
been observed to function correctly. Any defective item
must be returned in its original packing.

PRELIMINARY EXAMINATION

FUNCTIONAL DESCRIPTION

SPECIFICATIONS

1) BNC connector for pH (or ORP) electrode.
2) Liquid Crystal Display.
3)

mV

 key, to display the mV (ORP) readings when using

an ORP electrode or the mV equivalent to pH values
when using a pH electrode.

4)

T

rimmer

 

for pH OFFSET calibration.

5)

T

rimmer for pH SLOPE calibration.

6)

°C

 key, to display the temperature measurement.

7)

pH

 key, to display the pH value.

8)

ON/OFF 

key.

9) Socket for 

HI 7669AW

 temperature probe.

Range

0.00 to 14.00 pH
±1999 mV
0.0 to 100.0°C

Resolution

0.01 pH / 1 mV / 0.1°C

Accuracy 

(@20°C/68°F)

±0.01 pH / ±1 mV
±0.4°C

Typical EMC Deviation

±0.07 pH / ±5 mV / ±1°C

pH Calibration

Manual, at 2 points through
offset and slope trimmers

Offset Calibration

±1 pH

Slope Calibration

from 85 to 105%

ORP Calibration

Factory calibrated

Temperature Calibration

Factory calibrated

Temperature Compensation

Automatic, from 0 to 70°C

Probes 

(

included

)

HI 1230B

 pH-electrode

HI 7669AW

 temperature probe

Battery Type

1 x 9V alkaline

Battery Life

100 hours of continuous use

Environment

0 to 50°C (32 to 122°F);
max 95% RH non-condensing

Dimensions

164 x 76 x 45 mm
(6.5 x 3.0 x 1.8")

Weight

180 g (6.3 oz.)

ACCESSORIES

HI 1230B

Plastic body, double-junction, gel, combination pH-
electrode with BNC connector and 1 m cable

HI 3131B

Glass-body, refillable, combination platinum ORP-
electrode with BNC connector and 1 m cable

HI 7669AW

Temperature probe with 1 m screened cable

HI 7004M

pH 4.01 buffer solution, 230 ml bottle

HI 7006M

pH 6.86 buffer solution, 230 ml bottle

HI 7007M

pH 7.01 buffer solution, 230 ml bottle

HI 7009M

pH 9.18 buffer solution, 230 ml bottle

HI 7010M

pH 10.01 buffer solution, 230 ml bottle

HI 70300M

Storage solution, 230 ml bottle

HI 7061M

Cleaning solution, 230 ml bottle

HI 7091M

Reducing pretreatment solution, 230 ml bottle

HI 7092M

Oxidizing pretreatment solution, 230 ml bottle

HI 731326

Calibration screwdriver (20 pcs)

HI 76405

Electrode holder

Consult the Hanna General Catalog for a complete and
wide selection of electrodes.

Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.

Recommendations for Users

Before using this product, make sure that it is entirely suitable for the environment in which it
is used. Operation of this instrument in residential area could cause unacceptable interference to
radio and TV equipment, requiring the operator to take all necessary steps to correct interference.

The glass bulb at the end of the electrode is sensitive to electrostatic discharges. Avoid
touching this glass bulb at all times.

To maintain the EMC performance of equipment, the recommended cables noted in the user's
manual must be used. Any variation introduced by the user to the supplied equipment may
degrade the instruments' EMC performance.

To avoid electrical shock, do not use these instruments when voltages at the measurement surface
exceed 24Vac or 60Vdc. To avoid damage or burns, do not perform any measurement in microwave
ovens.