40
Make sure the instrument has been calibrated before taking DO measurements.
DIRECT MEASUREMENT
To measure the DO of a sample using the Direct reading
mode:
• Press
and then
to select DO measure
mode.
• Select the
Direct
reading mode (see DO Setup).
• Submerge the DO probe and allow time for the reading
to stabilize.
• The measured DO value will be displayed together with
the temperature and pressure values.
Notes: For accurate DO measurements, a water movement of 0.3 m/s is required. This is
to ensure that the oxygen-depleted membrane surface is constantly replenished. A moving
stream will provide adequate circulation.
If the reading is out of range, “-----” will be displayed.
DIRECT/AUTOHOLD MEASUREMENT
To measure DO of a sample using the Direct/AutoHold
reading mode:
• Select the
Direct/AutoHold
reading mode (see DO
Setup).
• If pressing
, the “AutoHold” indicator will start
blinking on the display until the stability criterion is
reached. The DO value will be frozen on the display,
along with “AutoHold” indicator.
• To return to normal measure mode press
.
DO MEASUREMENT