Clear Calibration

This function only works for the 

Percent Scale.

To clear calibration:

• Press 

 while in 

Salinity mode.

• Press 

.

• Use 

 or 

 to highlight the 

Clear Calibration option.

• Press 

 to clear calibration. A pop-up menu will be displayed asking for confirmation (if calibration is

available).

• Press 

 to confirm or press 

 to cancel operation.

Salinity Scale

Note:

See 

Salinity Measurement for a description of these scales.

The meter has three ocean salinity scales: 

Natural Sea Water 1966, Practical Scale 1978, Percent Scale [%].

To select the desired salinity measurement scale:

• Press 

 while in 

Salinity measure mode.

• Press 

.

• Use 

 or 

 to highlight the 

Salinity Scale option.

• Press 

 and use 

 or 

 to highlight the

desired option.

• Press 

 to confirm your selection or press 

 to

cancel operation.

Sample ID

 

see 

Conductivity Setup section.

Log 

see 

Conductivity Setup section.

Alarm

 

see 

Conductivity Setup section.

Содержание HI 5321-01

Страница 1: ...1 HI 5321 Conductivity Resistivity TDS Salinity Temperature Bench Meter Instruction Manual www hannainst com...

Страница 2: ...s first obtain a Returned Goods Authorization number from the Technical Service department and then send it with shipping costs prepaid When shipping any instrument make sure it is properly packed for...

Страница 3: ...P 14 CONDUCTIVITY SETUP 21 RESISTIVITY SETUP 35 TDS SETUP 37 SALINITY SETUP 39 CONDUCTIVITY CALIBRATION 41 CONDUCTIVITY MEASUREMENT 43 USP EVALUATION 44 RESISTIVITY MEASUREMENT 47 TDS MEASUREMENT 48 S...

Страница 4: ...Conductivity Probe with built in temperature sensor and ID HI 76404W Electrode Holder Conductivity Calibration Solutions Kit 12 Vdc Power Adapter Instruction Manual HI 5321 01 is supplied with 12 Vdc...

Страница 5: ...n up to four points probe offset calibration Single fixed point salinity calibration Percent Scale only AutoHold feature to freeze first stable reading on the LCD Two selectable alarm limits for condu...

Страница 6: ...5321 DESCRIPTION FRONT PANEL REAR PANEL 1 Liquid Crystal Display LCD 2 Capacitive touch keypad 3 ON OFF switch 4 Power adapter socket 5 Conductivity probe connector 6 USB connector FUNCTIONAL DESCRIP...

Страница 7: ...on depending on the current menu e g and in Measure mode LCD GENERAL DESCRIPTION To enter exit calibration mode To select the desired measurement mode Conductivity Resistivity TDS Salinity To enter Se...

Страница 8: ...r b i l a c C E m c S m 8 1 1 1 m c S m 0 0 0 8 m c S m 8 8 2 1 m c S m 0 0 0 5 m c S m 3 1 4 1 m c S 0 0 4 8 n o i t i n g o c e r e b o r p y t i v i t c u d n o C s e Y n o i t a s n e p m o c e r...

Страница 9: ...5 2 n o i t u l o s e R K 1 0 F 1 0 C 1 0 y c a r u c c A e b o r p t u o h t i w K 2 0 F 4 0 C 2 0 n o i t a r b i l a C C 0 0 1 0 5 0 s t n i o p 3 n i n o i t a r b i l a c r e s U l e n n a h c t...

Страница 10: ...covered by hand or other objects at the meter power on Turn the instrument on from the power button located on the rear panel of the instrument Please wait until the instrument finishes the initializa...

Страница 11: ...Reminder messages area Press The instrument will display the detailed GLP data For measurement mode s the following display configurations are available Basic Graph and Log History A GLP display is a...

Страница 12: ...rea Press Press to begin online graph To Zoom Graph Press then and will appear in virtual keys Use and to move along the X Time axis of the graph Press to access the zoom menu for Y axis Use or for zo...

Страница 13: ...while in Measure mode The Choose Display Configuration message will be displayed in the Reminder messages area Press The instrument will display the log history regarding the selected Measure mode Not...

Страница 14: ...LCD To access a System Setup option Use or to highlight the desired option Press to access the selected option The following is a detailed description of the System Setup option screen Beeper This opt...

Страница 15: ...ation GLP Data Use this option to customize log GLP information with specific identification data When enabled these ID tags will be included in the GLP section of all data logs Each data field can us...

Страница 16: ...without saving or to return to the editing mode Otherwise the modified options are saved automatically Date Time Set the current date time and the format in which they appear These parameters will be...

Страница 17: ...r on the LCD notifying the user that a date time conflict has occured and some time dependent modes could work improperly e g Measure GLP Log LCD Setup This option allows the user to set the Contrast...

Страница 18: ...lor Palette This option allows the user to choose a desired color palette To select the Color Palette Press while in Measure mode Press Use or to select the Color Palette option Press and use or to hi...

Страница 19: ...arest Hanna Customer Service for factory calibration To view the Meter Information Press while in Measure mode Press Use or to select the Meter Information option Press to acces the Meter Information...

Страница 20: ...user to erase all user settings and reset the instrument to the default factory settings To restore the Factory Settings Press while in Measure mode Press Use or to select the Restore Factory Settings...

Страница 21: ...Profile can be named and recalled at a moment s notice A profile is a sensor setup complete with measurement units logging and display preferences calibration standards Standards including custom setu...

Страница 22: ...by using and to highlight the desired character and then press to add it to the text bar It is also possible to delete the last character by positioning the cursor on the Backspace character and pres...

Страница 23: ...screen Use or to select the desired profile and press Press to return to the previous menu Reading Mode This option allows the user to select between Direct Direct AutoHold or Direct USP conductivity...

Страница 24: ...tion The user can choose from the following options Linear the meter will automatically compensate the conductivity using the following formula where Cref conductivity at reference temperature CI cond...

Страница 25: ...nce temperature of 25 C To set the reference temperature Use or to highlight the Temperature option Press and then use or to highlight the Reference Temperature option Press and then use or to increas...

Страница 26: ...ess to cancel operation Calibration Cell Constant The conductivity probe can be calibrated using the conductivity standards or by entering the cell constant of the probe by the user Using standard sol...

Страница 27: ...calibration Press to choose Single Point calibration The following options are available for calibration Standard Recognition The user can choose between Automatic recognition from 6 Hanna standards...

Страница 28: ...ss to cancel operation Set Reminder Period Daily reminder the user can set the time of day when the reminder is to appear Periodic reminder the user can set the time from the last calibration days hou...

Страница 29: ...escape without saving and return to the Calibration options Cell constant manual editing The conductivity probe can also be calibrated by entering the cell constant value To edit the cell constant val...

Страница 30: ...eters are available ID Increment mode and Edit Sample ID ID Increment Choose None to identify a sample with a text tag Choose Automatic to identify a sample with a numeric tag This number will be incr...

Страница 31: ...It is also possible to delete the last character by positioning thecursorontheBackspacecharacter andpressing Press to return to Sample ID option If the Saving Confirmation is enabled press to accept t...

Страница 32: ...nductivity mode Press Use or to highlight the Log option Press and use or to highlight the Logging Type option Press and use or to highlight the desired option Press to confirm your selection or press...

Страница 33: ...g the user that a new lot can be created only if the Logging Type is set as Manual To generate a New Lot Press while in Conductivity mode Press Use or to select the Log option Press and use or to high...

Страница 34: ...Press and use or highlight the Alarm State option Press and use or to highlight the desired option Press to confirm your selection or press to cancel operation Alarm Limits This option allows the use...

Страница 35: ...p option Use or to select the desired option Press to confirm your selection The following is a description of the Resistivity Setup option screens Profile see Conductivity Setup section RESISTIVITY S...

Страница 36: ...o select the units Press while in Resistivity mode Press Use or to highlight the Units option Press to confirm and then use or to highlight the desired unit Press to confirm or press to cancel operati...

Страница 37: ...ected option The following is a description of the TDS Setup option screens Profile see Conductivity Setup section Reading Mode see Resistivity Setup section Temperature see Conductivity Setup section...

Страница 38: ...s e g fertilizers is 0 70 Example TDS factor 0 5 S cm x 0 41 0 205 ppm NaCl The default value is 0 50 This option allows the user to set the TDS factor Press while in TDS mode Press Use or to highligh...

Страница 39: ...Profile see Conductivity Setup section Reading Mode see Resistivity Setup section Temperature see Conductivity Setup section To set one of the Temperature options Press while in Salinity mode Press U...

Страница 40: ...ity Scale Note See Salinity Measurement for a description of these scales The meter has three ocean salinity scales Natural Sea Water 1966 Practical Scale 1978 Percent Scale To select the desired sali...

Страница 41: ...ty and Natural Sea Water and Practical Sea Water Salinity readings are automatically derived from the conductivity readings so conductivity calibration is required OFFSET CALIBRATION The meter allows...

Страница 42: ...mpt for the custom standard value Press to finish the calibration or to abort calibration The probe should be rinsed in deionized water Shake off excess water Note The calculated cell constant will be...

Страница 43: ...Clear all old calibrated standards Make sure the instrument has been calibrated before taking conductivity measurements DIRECT MEASUREMENT To measure the conductivity of a sample using the Direct read...

Страница 44: ...onductivity probe has been calibrated in conductivity standards in the lowest measurement range Press and then select the desired USP stage In this measure mode the user can check for water quality us...

Страница 45: ...C Look up the corresponding conductivity value in the table on the next page If the measured conductivity is lower than the conductivity in the table then the water meets the USP requirements Otherwi...

Страница 46: ...ults This may be printed using HI 92000 software Note A separate pH meter and sensor is required Stage 3 this is an off line test that studies the pH and CO2 If the water sample has failed Stage 1 and...

Страница 47: ...ample from the stage 2 test and increase its ionic strength for a pH measurement at 25 C Use 100 mL of Stage 2 water and add 300 L saturated KCl to the sample Calibrate a pH sensor in pH 4 01 and pH 6...

Страница 48: ...the same as for the conductivity measurement see Conductivity Measurement section DIRECT AUTOHOLD MEASUREMENT To measure resistivity of a sample using the Direct AutoHold reading mode Select the Dire...

Страница 49: ...eeve Position probe away from the wall or bottom of the container Enter in calibration mode by pressing Wait for measurement to stabilize Press to finish salinity calibration or press to cancel calibr...

Страница 50: ...e following coefficients and k 0 0162 and Seawater temperature coefficient rT c0 c1 T c2 T2 c3 T3 c4 T4 S a0 a1 RT 1 2 a2 RT a3 RT 3 2 a4 RT 2 a5 RT 5 2 T 15 b0 b1 RT 1 2 b2 RT b3 RT 3 2 b4 RT 2 b5 RT...

Страница 51: ...e EC channel and temperature pH channel Press to start the temperature calibration Adjust the temperature preset value using or when necessary Insert the EC probe into the beaker with water at 0 C Wai...

Страница 52: ...2 Direct Reading Mode and Manual Log Real time continuous measurements are on display and snapshots of measurement data are stored in the Manual log when the user presses Subsequent snapshots will be...

Страница 53: ...marked with an H 5 Direct AutoHold Reading Mode and Auto Hold Log Press and then keys to initiate and automate the capture of stable data which is stored in the Recall Manual Log file During the proc...

Страница 54: ...together with GLP information last calibration date and calibrated standards if a calibration has been performed on the selected mode and the logged values measured value temperature value temperatur...

Страница 55: ...Logged lots should also be deleted whenever Limited Automatic Logging Space or Automatic Log Is Full message appears on the LCD in the Reminder messages area Data transmission from the instrument to...

Страница 56: ...be to ensure representative sample is measured within the sleeve of the probe If required wait for the probe to reach thermal equilibrium with the sample PERIODIC MAINTENANCE Inspect the probe and the...

Страница 57: ...m e h t n o i t u l o s n i s i e b o r p y f i r e V t n s e o d t n e m u r t s n i e h T e r u t a r e p m e t e h t e r u s a e m e b o r p e h t m o r f e r u t a r e p m e t e b o r p e h T e h...

Страница 58: ...25 pcs HI 7030M 12880 S cm 230 mL bottle HI 7030L 12880 S cm 500 mL bottle HI 8030L 12880 S cm 500 mL FDA approved bottle HI 7034M 80000 S cm 230 mL bottle HI 7034L 80000 S cm 500 mL bottle HI 8034L 8...

Страница 59: ...follow all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid electrical shock do not use the...

Страница 60: ...Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Technical Support for Customers Tel 800 426 6287 Fax 401 765 7575 E mail tech hannainst com www hannainst com Local Sales and...

Отзывы: