Alkaline Error

TEMPERATURE-RESISTANCE CORRELATION FOR

HANNA pH SENSITIVE GLASS

The resistance of glass electrodes partially depends on the temperature.
The lower the temperature, the higher the resistance. It takes longer time
for the reading to stabilize if the resistance is higher. In addition, the
response time will suffer to a greater degree at temperatures below 10°C.

Since the resistance of the pH electrode is in the range of 200 Mohm,
the current across the membrane is in the pico-Ampere range. Large
currents can disturb the calibration of the electrode for many hours.
For these reasons 

high humidity environments, short circuits and

static discharges

 are detrimental to a stable pH reading.

The pH electrode's life also depends on the temperature. If constantly
used at high temperatures, the electrode life is drastically reduced.

Typical Electrode Life
Ambient  Temperature

1- 3 years

90 °C

Less than 4 months

120°C

Less than 1 month

High concentrations of sodium ions interfere with readings in alkaline
solutions; the pH at which the interference starts to be significant
depends upon the composition of the glass. This interference is the
alkaline error and causes the pH to be underestimated. Hanna's
glass formulations have the indicated characteristics.

-20 -20

-10

0 +10+20 +30+40 +50+60+70 +80+90

° C

-10

-10

-10

-10

-10

-10

 Ω

2x10

9

1x10

9

2x10

8

1x10

8

2x10

7

1x10

7

-20   -10      0 

 +

10 

 +

20

+

30

+

40 

+

50 

+

60 

 +

70 

+

80 

 +

90 

   °C

1.0 Mol L

-1

 Na

+

0.1 Mol L

-1

 Na

+

Sodium Ion Correction for the Glass at 20-25°C
Concentration

pH

  Error

13.00
13.50
14.00
12.50
13.00
13.50
14.00

0.10
0.14
0.20
0.10
0.18
0.29
0.40

Note:

 To prevent damage to the electrode, remove the pH electrode

from the sample before turning the meter off.

If the meter is OFF, detach the electrode from the meter before
immersing the electrode in the storage solution.

PREPARATION

Remove the protective cap.
DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT.
This is normal with electrodes and they will disappear when rinsed
with water.
During transport tiny bubbles of air may form inside the glass bulb.
The electrode cannot function properly under these conditions. These
bubbles can be removed by "shaking down" the electrode as you
would do with a glass thermometer.
If the bulb and/or junction are dry, soak the electrode in 

HI 70300

Storage Solution for at least one hour.

For refillable electrodes:

If  the fill solution (electrolyte) is less than 1 cm (½") below the fill
hole, add 

HI 7082 3,5M KCl Electrolyte Solution

 for double

junction or 

HI 7071 3,5M KCl+AgCl Electrolyte Solution

 for single

junction electrodes.
For a faster response unscrew the fill hole screw during measurements.

For AmpHel electrodes:

If the electrode does not respond to pH changes, the battery is run
down and the electrode should be replaced.

MEASUREMENT

Rinse the electrode tip with distilled water.
Immerse the tip (4cm /1½") in the sample and stir gently for
approximately 30 seconds.
For a faster response and to avoid cross contamination of the samples,
rinse the electrode tip with a few drops of the solution to be tested,
before taking measurements.

STORAGE

To minimize clogging and ensure a quick response time, the glass
bulb and the junction should be kept moist and not allowed to dry
out.

ELECTRODE CONDITIONING

AND MAINTENANCE

Содержание HI 504920

Страница 1: ...Instruction Manual HI 504920 Remote User Interface pH Meter w w w h a n n a i n s t c o m...

Страница 2: ...PRELIMINARY EXAMINATION GENERAL DESCRIPTION Dear Customer Thank you for choosing a Hanna Instruments Product Please read this instruction manual carefully before using the instru ment This manual wil...

Страница 3: ...Preventing System BEPS detects when the batter ies level becomes weak When the batteries are getting low a clear indication is displayed on LCD to warn the user of this condition However the meter co...

Страница 4: ...efore turning the meter off When the meter is OFF detach the electrode from the meter before immersing it in the storage solution SPECIFICATIONS Range pH 4 00 to 19 99 pH mV 600 0 mV 2000 mV autorangi...

Страница 5: ...When the CFM symbol blinks the read ing is stable and calibration can be confirmed 8 Press CFM to confirm the first buffer 9 If everything is satisfactory the LCD will display Stor and then the expect...

Страница 6: ...e above procedure only when readings are stable 7 To exit the calibration mode and return to normal measurement press TXCAL mV CALIBRATION Contact your nearest Hanna Service Center for periodic mV cal...

Страница 7: ...nterface cable connected between HI 504920 and HI 504910 step 1 it can happen that the HI 504910 must be located in a place difficult to reach when the HI 504920 pH probe is immersed in the solution I...

Страница 8: ...0 Calibration alarm time out 01 to 99 days OFF OFF 20 Auto Off Power down time out ON OFF ON 21 Firmware version 22 Battery level test 90 HI 504910 terminal mode 99 Password2 0000 to 9999 0000 Setup c...

Страница 9: ...y checks for entered time date accuracy as follows 0 hh 23 0 mm 59 01 dd 28 29 30 31 1 MM 12 1998 YYYY 2097 2 To change the password the correct code must be entered first If the password has been for...

Страница 10: ...When turned on the meter checks if the time out time has expired If yes the message Cal date scrolls across the LCD and the DATE tag blinks as a reminder LAST CALIBRATION DATA Last calibration data ar...

Страница 11: ...ers In this case check the correct polarity of the adapter before connecting it to the meter BATTERY REPLACEMENT The LCD will then display the last pH calibration date Press DISPLAY to scan remaining...

Страница 12: ...4 00 12 50 13 00 13 50 14 00 0 10 0 14 0 20 0 10 0 18 0 29 0 40 Note To prevent damage to the electrode remove the pH electrode from the sample before turning the meter off If the meter is OFF detach...

Страница 13: ...torage Solution Follow the Preparation Procedure above before taking measurements Note NEVER STORE THE ELECTRODE IN DISTILLED WATER OR DRY PERIODIC MAINTENANCE Inspect the electrode and the cable The...

Страница 14: ...SOLUTIONS HI 7091L Reducing Pretreatment solution 500 mL bottle HI 7092L Oxidizing Pretreatment solution 500 mL bottle ELECTRODES HI 62920 pH electrode with built in temperature sensor matching pin Ti...

Страница 15: ...2 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 01...

Отзывы: