CONDUCTIVITY CALIBRATION
CONDUCTIVITY CALIBRATION
CONDUCTIVITY CALIBRATION
CONDUCTIVITY CALIBRATION
CONDUCTIVITY CALIBRATION
It is recommended to calibrate the instrument frequently, especially if high accuracy is required.
The conductivity range should be recalibrated:
• Whenever the conductivity probe is replaced.
• At least once a week.
• Before USP measurements.
• After testing aggressive chemicals.
• When calibration reminder is activated (“Conductivity Cal Expired”).
• If the readings are far from the calibration point.
Note:
TDS and Resistivity readings are automatically derived from the conductivity readings and no specific
calibration is needed.
OFFSET CALIBRATION
The meter allows the user to calibrate the probe for the offset.
• Set the meter for conductivity range;
• Select the automatic standard recognition (see
Conductivity Setup -> Calibration);
• Leave the dry probe in the air;
• Enter in calibration mode by pressing
;
• Wait to stabilize. The 0.000
µ
S calibration point will appear on the screen;
• Press
to finish the probe offset calibration.
• Press
to exit calibration mode or continue calibration in the other standard solutions.
Note:
The offset calibration can be performed only if it is perfomed first (no other calibration points present).
Clear the old calibration if it is present to calibrate for the offset.
CELL CONSTANT CALIBRATION (in solution)
Single Point Calibration
• Select the single point calibration (see
Conductivity Setup -> Calibration);
•
Pour a small quantity of the standard solution into a clean beaker. If possible, use plastic beakers to
minimize any EMC interferences.
• For accurate calibration and to minimize cross-contamination, use two beakers for each standard solution
(one for rinsing the probe and another one for calibration).