31
Log
This option allows the user to edit the settings related to the logging feature, as following:
Logging Type
Three logging types are available:
Automatic, Manual and AutoHold.
Automatic logging - the readings are logged automatically at constant
time intervals (see
Sampling Period option).
Manual logging ( log on demand)- the readings are logged each time
is pressed.
AutoHold logging - the readings are logged automatically at each
auto-hold event occured.
To set the sample logging type:
• Use
or
to highlight the
Logging Type option.
• Press
and use
or
to choose from
Automatic, Manual or Auto Hold.
• Press
to confirm your selection or press
to
cancel operation.
Logging Data Configuration
This option allows the user to select the parameters that accompany a
logged value:
Date/Time, Calibration Data, Sample ID, Instrument ID,
Operator ID, Company Name, Aditional Info 1 and Aditional Info 2.
To customise the logging data configuration:
• Use
or
to highlight the
Logging Data
Configuration option.
• Press
and then use
or
to enable
the parameter by selecting
Yes or to disable it by selecting No for
each option.
• Press
to return to the previous menu.