CLEANING PROCEDURE
CLEANING PROCEDURE
CLEANING PROCEDURE
CLEANING PROCEDURE
CLEANING PROCEDURE
• General
Soak in Hanna
HI 7061
or
HI 8061
General Cleaning
Solution for approximately ½ hour.
• Protein
Soak in Hanna
HI 7073
or
HI 8073
Protein Cleaning
Solution for 15 minutes.
• Inorganic
Soak in Hanna
HI 7074
Inorganic Cleaning Solution
for 15 minutes.
• Oil/grease
Rinse with Hanna
HI 7077
or
HI 8077
Oil and Fat
Cleaning Solution.
IMPORTANT:
After performing any of the cleaning procedures, rinse the
electrode thoroughly with distilled water, refill the reference chamber
with fresh electrolyte (not necessary for gel-filled electrodes) and soak the
electrode in
HI 70300
or
HI 80300
Storage Solution for at least 1 hour