• When the reading is stable and close to the selected buffer, the
“
CFM
” tag will blink and if enabled, an audible signal will sound.
• Press the
CFM
key to confirm the calibration. The calibrated value
will be displayed on the primary LCD and the second expected buffer
value on the secondary LCD.
• After the first calibration point is confirmed, submerse the pH electrode
and the temperature probe approximately 3 cm (1¼”) into the
second buffer solution and stir gently. The temperature probe should
be close to the pH electrode.
• If necessary, press the
ARROW
keys to select a different buffer value.
Note:
The instrument will automatically skip the buffer used for the first
point. It also skips 6.86 if 7.01 was used, and vice versa.
Likewise, it will skip 9.18 if 10.01 has been used, and vice versa.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer the “
CFM
”
tag will blink.
• Press
CFM
to confirm calibration.
• The calibrated value is then displayed on the
primary LCD and and the third expected buffer
value on the secondary LCD.