23
• Press
CAL
when the instrument is in RELATIVE mV measurement
mode. The “
mV
” and “ ” tags will blink. Absolute mV is displayed
on the primary LCD and “
AbS
” message is displayed on the secondary
LCD.
• When the absolute reading is stable and in measurement range, the
instrument asks for confirmation.
• If the reading is out of range, “
WRONG
” tag will be displayed.
• Press
CFM
to confirm the absolute value. The instrument will display
“
0.0 mV
” on the primary LCD and “
rEL
” message on the secondary
LCD. In this moment the relative mV offset is equal to absolute mV
reading.
• Use the
ARROW
keys if you want to change the displayed relative
mV value.
• Press
CFM
to confirm the relative mV value. The relative mV offset is
displayed on the primary LCD and “
OFF
” message on the secondary
LCD for a few seconds. The instrument returns to measurement mode.
Note
: The relative mV value can be changed only inside the relative mV
offset window (± 2000 mV).