18
SECOND CUSTOM BUFFER
Press
CAL
when
“cb2”
is displayed. The custom buffer (disabled –
“no”
or “
0
” to “
16
” pH) and “
CFM
” will start blinking.
Press the
ARROW
keys to change the custom buffer value.
Press
CFM
to save the modified custom buffer value.
Press
CAL
to escape without saving.
ONE-POINT CALIBRATION BEHAVIOR
Press
CAL
when “
1 Pnt
” message is displayed on the secondary LCD.
One of the two options (“
Pnt
” or “
OFFS
”) and “
CFM
” will start
blinking (see pH CALIBRATION PROCEDURE for details, page 8).
Press the
ARROW
keys to toggle between “
Pnt
” and “
OFFS
” options.
Press
CFM
to save the behavior for one-point calibration.
Press
CAL
to escape without saving.
CURRENT TIME
Press
CAL
when the current time is displayed. The hour and “
CFM
” will
start blinking.
Press the
ARROW
keys to change the hour.
Press
RANGE
(
HI 2213
) or
u
(
HI 2212
). The minutes and “
CFM
” will
start blinking.
Press the
ARROW
keys to change the minutes.
Press
CFM
to save the modified value.
Press
CAL
to escape without saving.