background image

20

21

ACCESSORIES

ACCESSORIES

ACCESSORIES

ACCESSORIES

ACCESSORIES

pH BUFFER SOLUTIONS

HI 70004P

pH 4.01 Buffer Sachets, 20 mL, 25 pcs

HI 70007P

pH 7.01 Buffer Sachets, 20 mL, 25 pcs

HI 70010P

pH 10.01 Buffer Sachets, 20 mL, 25 pcs

HI 7004L

pH 4.01 Buffer Solution, 500 mL bottle

HI 7006L

pH 6.86 Buffer Solution, 500 mL bottle

HI 7007L

pH 7.01 Buffer Solution, 500 mL bottle

HI 7009L

pH 9.18 Buffer Solution, 500 mL bottle

HI 7010L

pH 10.01 Buffer Solution, 500 mL bottle

HI 8004L

pH 4.01 Buffer Sol. in FDA approved bottle, 500 mL

HI 8006L

pH 6.86 Buffer Sol. in FDA approved bottle, 500 mL

HI 8007L

pH 7.01 Buffer Sol. in FDA approved bottle, 500 mL

HI 8009L

pH 9.18 Buffer Sol. in FDA approved bottle, 500 mL

HI 8010L

pH 10.01 Buffer Sol. in FDA approved bottle, 500 mL

ELECTRODE STORAGE SOLUTIONS

HI 70300L

Storage Solution, 460 mL bottle

HI 80300L

Storage Solution in FDA approved bottle, 460 mL

ELECTRODE CLEANING SOLUTIONS

HI 70000P

Electrode Rinse Sachets, 20 mL, 25 pcs

HI 7061L

General Cleaning Solution, 460 mL bottle

HI 7073L

Protein Cleaning Solution, 460 mL bottle

HI 7074L

Inorganic Cleaning Solution, 460 mL bottle

HI 7077L

Oil & Fat Cleaning Solution, 460 mL bottle

HI 8061L

General Cleaning Sol. in FDA approved bottle, 460 mL

HI 8073L

Protein Cleaning Solution in FDA approved bottle, 460 mL

HI 8077L

Oil & Fat Cleaning Sol. in FDA approved bottle, 460 mL

ELECTRODE REFILL ELECTROLYTE SOLUTIONS

HI 7071

3.5M KCl+AgCl Electrolyte, 4x50 mL, for single junction
electrodes

HI 7072

1M KNO

3

 Electrolyte, 4x50 mL

HI 7082

3.5M KCl Electrolyte, 4x50 mL, for double junction
electrodes

HI 8071

3.5M KCl+AgCl Electrolyte in FDA approved bottle,
4x50 mL, for single junction electrodes

HI 8072

1M KNO

3

 Electrolyte in FDA approved bottle, 4x50 mL

HI 8082

3.5M KCl Electrolyte in FDA approved bottle, 4x50 mL

Typical Electrode Life

Ambient  Temperature

1- 3 years

90 °C

Less than 4 months

120 °C

Less than 1 month

Alkaline Error

High concentrations of sodium ions interfere with readings in alkaline

solutions. The pH at which the interference starts to be significant

depends upon the composition of the glass. This interference is called

alkaline error and causes the pH to be underestimated. Hanna’s glass

formulations have the indicated characteristics.

1.0 Mol L

-1

 Na

+

0.1 Mol L

-1

 Na

+

Sodium Ion Correction for Glass at 20-25 °C

Concentration

pH

   Error

13.00

13.50

14.00

12.50

13.00

13.50

14.00

0.10

0.14

0.20

0.10

0.18

0.29

0.40

The resistance of glass electrodes partially depends on the temperature.

The lower the temperature, the higher the resistance. It takes more time

for the reading to stabilize if the resistance is higher. In addition, the

response time will suffer to a greater degree at temperatures below 25 °C.

Since the resistance of the pH electrode is in the range of 50-200 Mohms,

the current across the membrane is in the pico Ampere range. Large

currents can disturb the calibration of the electrode for many hours.
For these reasons high humidity environments, short circuits and

static discharges are detrimental to a stable pH reading.
The pH electrode’s life also depends on the temperature. If constantly

used at high temperatures, the electrode life is drastically reduced.

TEMPERATURE CORRELATION

TEMPERATURE CORRELATION

TEMPERATURE CORRELATION

TEMPERATURE CORRELATION

TEMPERATURE CORRELATION

FOR 

FOR 

FOR 

FOR 

FOR  pppppH SENSITIVE GLASS

H SENSITIVE GLASS

H SENSITIVE GLASS

H SENSITIVE GLASS

H SENSITIVE GLASS

Содержание HI 2211

Страница 1: ...1 Instruction Manual Microprocessor based pH mV C Bench Meters HI 2210 HI 2211 www hannainst com...

Страница 2: ...pping any instrument make sure it is properly packed for complete protection WARRANTY 3 PRELIMINARY EXAMINATION 4 GENERAL DESCRIPTION 4 FUNCTIONAL DESCRIPTION 5 SPECIFICATIONS 6 OPERATIONAL GUIDE 7 pH...

Страница 3: ...mV and temperature simultaneously together with graphic symbols The calibration process is guided step by step through clear indications on the LCD A stability indicator makes the calibration procedur...

Страница 4: ...probe approximately 4 cm 1 into the sample to be tested and stir gently Allow time for the electrode to stabilize The pH is displayed on the primary LCD and the temperature on the secondary LCD If the...

Страница 5: ...temperature probe into the sample as close as possible to the electrode and wait for a few seconds If the temperature of the sample is known manual temperature compensation can be performed by disconn...

Страница 6: ...in this situation CAL 2 BUF pH BUFFER pH p p p p pH CALIBRATION H CALIBRATION H CALIBRATION H CALIBRATION H CALIBRATION Calibrate the instrument often especially if high accuracy is required The inst...

Страница 7: ...down the CAL MEM keys then power on the instrument The CAL message will appear and the secondary LCD will show 0 0 C Immerse the temperature probe in the vessel with ice and water as near as possible...

Страница 8: ...9 3 0 4 4 8 6 9 9 6 1 1 9 2 9 9 0 4 4 0 1 4 0 4 4 8 6 8 9 6 7 0 9 8 8 9 5 4 3 1 1 5 0 4 3 8 6 8 9 6 4 0 9 5 8 9 0 5 2 2 1 6 0 4 3 8 6 8 9 6 1 0 9 2 8 9 5 5 1 3 1 8 0 4 4 8 6 8 9 6 9 9 8 9 7 9 0 6 0 4...

Страница 9: ...e Preparation Procedure on page 16 before taking measurements Note NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED WATER PERIODIC MAINTENANCE PERIODIC MAINTENANCE PERIODIC MAINTENANCE PERIODIC MAI...

Страница 10: ...s e g n a r e t y l o r t c e l e e h t k c e h C e l a r e n e g e h t d n a l e v e l e d o r t c e l e e h t f o e t a t s s w o h s y a l p s i d e h T d n a V m e h t n i e g n a r f o t u O e l...

Страница 11: ...n FDA approved bottle 4x50 mL for single junction electrodes HI 8072 1M KNO3 Electrolyte in FDA approved bottle 4x50 mL HI 8082 3 5M KCl Electrolyte in FDA approved bottle 4x50 mL Typical Electrode Li...

Страница 12: ...IA 0 29 S VERSION HI 1331 ORP PRETREATMENT SOLUTIONS HI 7091L Reducing Pretreatment Solution 460 mL bottle HI 7092L Oxidizing Pretreatment Solution 460 mL bottle pH ELECTRODES All electrodes part numb...

Страница 13: ...I 3230 S VERSION 120 mm 4 7 12 mm 0 5 HI 4430 S VERSION FC 100B Plastic body PVDF double junction refillable combination pH electrode Use general purpose for food industry FC 200B Plastic body PVDF op...

Страница 14: ...To avoid damage or burns do not perform any measurement in microwave ovens EXTENSION CABLE FOR SCREW TYPE ELECTRODES SCREW TO BNC ADAPTER HI 7855 1 Extension cable 1 m 3 3 long HI 7855 3 Extension cab...

Страница 15: ...2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 2739 3014 Fax 886 2 2739 2...

Отзывы: