![Hanna Instruments HI 22 Series Скачать руководство пользователя страница 5](http://html1.mh-extra.com/html/hanna-instruments/hi-22-series/hi-22-series_instruction-manual_3711544005.webp)
9
8
INSTALLATION
HI 21
and
HI 22
offer
a multitude of possibili-
ties, from single and
dual setpoints to ON/
OFF, proportional or
PID dosage, isolated
outputs with user-se-
lectable zoom,
bidirectional RS485, re-
corder outputs in
mAmps and Volts.
In addition,
HI 21
and
HI 22
are both
equipped with the ex-
clusive differential
input.
In a system with poor
grounding, it is possible
to have a ground cur-
rent flowing through
the reference junction.
This can cause a rapid
degradation of the elec-
trode. The Hanna
differential input re-
duces the likelihood of
ground loops.
See the diagram for a
recommended installa-
tion.
SPECIFICATIONS HI 21 & HI 22
Range
0.00 to 14.00 pH (
HI 21
Series only)
±2000 mV (
HI 22
Series only)
-9.9 to 120.0 °C
Resolution
0.01 pH (
HI 21
Series only)
1 mV (
HI 22
Series only)
0.1 °C
Accuracy
±0.02 pH (
HI 21
Series only)
(@20°C/68°F)
±2 mV (
HI 22
Series only)
±0.5 °C
Typical EMC Deviation
±0.05 pH (
HI 21
Series only)
±4 mV (
HI 22
Series only)
±1.0 °C
Analog Output
0 to 1 mA, 0 to 20 mA, 4 to 20 mA
0 to 5 Vdc, 1 to 5 Vdc, 0 to 10 Vdc
Digital Output
RS485
Power Supply Input
±5V (for amplified electrodes)
Power Supply
“-1” models: 115 ±10% Vac, 50/60 Hz
“-2” models: 230 ±10% Vac, 50/60 Hz
Power Consumption
15 VA
Over Current Protection
400 mA, 250V, FAST FUSE
Relays 1 and 2
Electromechanical relay SPDT contact outputs,
5A-250 Vac, 5A - 30 Vdc (resistive load)
(
HI 211YZ
,
HI 212YZ
and
HI 221YZ
)
Fuse protected: 5A, 250V FAST FUSE
Alarm Relay
Electromechanical relay SPDT contact output,
5A - 250 Vac, 5A - 30 Vdc (resistive load)
Fuse protected: 5A, 250V FAST FUSE
Environment
0 to 50°C (32 to 122°F);
RH max 85% non-condensing
Protection
IP 54
Dimensions
221 x 181 x 86 mm (8.7 x 7.1 x 3.4’’)
Weight
approximately 1.4 kg (3.1 lb.)