Hanna Instruments HI 22 Series Скачать руководство пользователя страница 21

41

40

PH BUFFER SELECTION (HI 21523 ONLY)

A one point pH calibration at a value different from standard
buffer is possible by directly entering the desired calibration
value.

• Pour a small quantity of the calibra-

tion solution in a beaker and then
press CAL to enter the calibration
mode.

• After the correct password is entered,

the control action stops and the pri-
mary LCD will display the pH value
using the current offset and slope,
with the  "CAL" and "

BUF

1

" indicators

lit and the probe indicator "   " blink-
ing.

• Press SETUP and the pH value on

the secondary LCD will start blink-
ing.

• By the  , 

and 

keys select the pH calibration value

of the buffer.

• Immerse the electrodes in the buffer solution.

• When the reading is stable press CFM to confirm the cali-

bration.

Note

During calibration with pH buffer solutions different from stan-
dard, the stability check function is not active. No “CAL” or
"

BUF

1

" indicators will blink, hence wait for the reading to stabi-

lize after having immersed the electrode in the calibration
solution.

Note

If the wrong password is entered the system reverts back and
restarts displaying the pH value.

Note

Calibration can be aborted by pressing SETUP before CFM .
Calibration is restarted with no changes to calibration data.

CALIBRATION WITH MANUAL TEMPERATURE COMPENSATION

• Enter the calibration procedure and press LCD to display

the temperature on the secondary LCD.

• Unplug any temperature probe that may

be attached to the meter. The "°C" sym-
bol will flash.

• Note the temperature of the buffer solutions with a

ChecktempC or another accurate thermometer with a reso-
lution of 0.1°C.

• Use   or   to manually adjust the display reading to the

value of the reference thermometer (e.g. 20°C).

Follow the calibration procedure above.

Note

To toggle between the pH buffer and
the temperature press LCD.

When a one-point calibration is carried out only the pH off-
set is computed and stored, while the pH slope is fixed
according to the theoretical values.

With a two-point calibration, offset and slope are computed
to fit the two calibration points. With a three-point calibra-
tion the offset and first slope values refers to pH 4.01 and
7.01 buffers, while the second slope refers to pH 7.01 and
10.01 buffers.

Note

If the process meter has never been cali-
brated or an EEPROM reset has occurred,
the meter continues to perform measure-
ment. However, the user is informed of a pH calibration
requirement by a blinking “CAL” (see “Start-up” section).

The device must be calibrated within the temperature range
of 

0

-

95°C

. Outside this range, the buffer pH values are not

reliable.

Содержание HI 22 Series

Страница 1: ...HI 21 HI 22 Series Wall mounted Microprocessor based pH and ORP Controllers Instruction Manual...

Страница 2: ...ment It will provide you with the necessary information for the correct use of the instrument as well as a precise idea of its versatility If you need additional technical information do not hesitate...

Страница 3: ...all packing materials until you are sure that the instru ment functions correctly Any damaged or defective items must be returned in their original packing materials together with the supplied accesso...

Страница 4: ...nd reverts back to normal mode in idle or control phases with the measure ment on the display In HI 21 series during pH calibration alternately displays pH buffer value or current temperature 3 CAL DA...

Страница 5: ...pH HI 21 Series only 1 mV HI 22 Series only 0 1 C Accuracy 0 02 pH HI 21 Series only 20 C 68 F 2 mV HI 22 Series only 0 5 C TypicalEMC Deviation 0 05 pH HI 21 Series only 4 mV HI 22 Series only 1 0 C...

Страница 6: ...s Power Supply Connect a 3 wire power cable to the terminal strip while paying attention to the correct line L earth PE and neutral N terminal connections Power 115VAC 100 mA 230VAC 50 mA Line Contact...

Страница 7: ...pH 1 pH no D1 25 to 4000 mV 50 mV 15 Relay 1 reset time 0 1 to 999 9 minutes 999 9mins no HI 21523 only 16 Relay 1 rate time 0 0 to 999 9 minutes 0 0 mins no HI 21523 only 21 Relay 2 mode M2 same as...

Страница 8: ...H2 S2 LA HA S1 D1 Code Valid Values Default PW 23 Relay 2 hysteresis H2 0 00 to 14 00 pH 1 pH no 0 to 4000 mV 50 mV 24 Relay 2 deviation D2 0 50 to 14 00 pH 1 pH no 25 to 4000 mV 50 mV 25 Relay 2 res...

Страница 9: ...nly high setpoint acid If M1 2 and M2 3 then S1 H1 S2 S1 LA HA S2 D2 If M1 4 and M2 1 then S1 S2 H2 S1 D1 LA HA S2 If M1 1 and M2 4 then S1 H1 S2 S2 D2 LA HA S1 If M1 3 and M2 4 then S1 S2 S2 D2 LA HA...

Страница 10: ...or base solution ON OFF 14 Setpoint Hysteresis Setpoint t0 t0 Tc t0 2Tc t0 3Tc dosage if available An upper boundary is imposed for acid base dosage time when relays are energized continuously i e whe...

Страница 11: ...ferent parameters i e the setpoint S1 or S2 the deviation D1 or D2 the reset time the rate time and the proportional control mode period Tc from 1 to 30 minutes Note User can disable the derivative an...

Страница 12: ...e the alarm function operates in a Normally Closed state and hence alarm is triggered if the wires are tripped or when the power is down 1 Starting from a solution with a pH or mV value different from...

Страница 13: ...ed to set off the alarm in abnormal circum stances forexample ifthedosingterminals are closed for too long a period In both cases the red LED s will also provide a visual warning signal The Fail Safe...

Страница 14: ...ground output and the second port to the current output or voltage output depending on which parameter is being used as depicted aside The type voltage or current and the range of the output analog si...

Страница 15: ...nts Data transmission from the instrument to the PC is possible with the HI 92500 Windows compatible application soft ware offered by Hanna Instruments The user friendly HI 92500 offers a variety of f...

Страница 16: ...re reading As additional feature the controller is also provided with two pins 5V and 0V in or der to apply the Fail Safe Open Line protection method To avoid erroneous readings in Open Line condition...

Страница 17: ...ading is 10 7 C the control action is active and no alarm condition is present Note If the controller is not in control or idle mode and the tem perature reading is requested through the TMR command t...

Страница 18: ...he calibration procedure Initial Preparation Pour small quantities of pH 7 01 HI 7007 and pH 4 01 HI 7004 and or pH 10 01 HI 7010 solutions into individual beakers If possible use plastic beakers to m...

Страница 19: ...ue For best accuracy however it is recommended that a two point calibration is performed HI 7007 CALIBRATION HI 7007 RINSE For accurate calibration use two beakers for each buffer solution the first o...

Страница 20: ...ration Two point Calibration Proceed as described above for one point cali bration usingpH 7 01asthefirstpoint butdo notquitcalibrationbypressingCALattheend Note The meter will automatically skip the...

Страница 21: ...ation is restarted with no changes to calibration data CALIBRATION WITH MANUAL TEMPERATURE COMPENSATION Enter the calibration procedure and press LCD to display the temperature on the secondary LCD Un...

Страница 22: ...t code 0 for mV calibration and press CFM to enter CAL will blink on the LCD until the meter confirms a steady reading OFFSET AND SLOPE DIRECT SELECTION HI 21523 ONLY Whenever the pH electrode offset...

Страница 23: ...lized at a point near the first calibration point CAL will stop blinking and an intermittent CFM icon will prompt the user to confirm the first calibration If the display stabilizes at a value signifi...

Страница 24: ...dc 3 0 Vdc 5 Vdc 1 5 Vdc 4 1 Vdc 5 Vdc 0 10Vdc 5 0 Vdc 10 Vdc Immerse the temperature probe in the second beaker as near to the Checktemp as possible and repeat the above procedure Calibration procedu...

Страница 25: ...heEEPROM Date Time Offset in mV for HI 21 only Up to two slopes for HI 21 only Up to three buffers While displaying this data the pH controller remains in con trol mode The procedure below indicates t...

Страница 26: ...data is composed of setup data and calibration data As for the setup data the calibration data is assigned default values when an EEPROM reset occurs An un calibrated meter can perform mea surement th...

Страница 27: ...together the LCD will look like this FAULT CONDITIONS AND SELFTEST PROCEDURES The fault conditions below may be detected by the software EEPROM data error I2C internal bus failure code dead loop EEPRO...

Страница 28: ...osition of squares Note A maximum of two keys may be pressed simultaneously to be properly recognized To exit the keyboard test procedure press LCD CAL and SETUP simultaneously EEPROM SELFTEST The EEP...

Страница 29: ...S Temperature has a significant effect on pH The calibration buffer solutions are effected by temperature changes to a lesser degree than normal solutions For manual temperature calibration please ref...

Страница 30: ...I 7082 for double junction electrodes if necessary Low Slope Refer to the cleaning procedure above If the bulb and or junction are dry soak the electrode in HI 70300 storage solution for at least one...

Страница 31: ...rbs hydrogen in the presence of reducing mediums This phenomenon is rough on the electrode To make correct redox measurements the following condi tions must prevail The surface of the electrode must b...

Страница 32: ...4 740 5 680 6 640 7 580 8 520 9 460 10 400 11 340 12 280 13 220 14 160 Reducing pretreatment immerse the electrode for a few min utes in HI 7091 Oxidizing pretreatment immerse the electrode for a few...

Страница 33: ...HI 3090B 5 BNC connector 5 m 16 5 cable double junction Pt glass body polymer filled RECOMMENDED pH ELECTRODES all electrodes are gel filled and with ce ramic junction unless otherwise indicated HI 1...

Страница 34: ...the electrode connector there is also a spade lug connection for the matching pin HI 3210B 5 BNC connector 5 m 16 5 cable double junction Pt plas tic body PVDF junction polymer filled GOLD ORP ELECTRO...

Страница 35: ...Goods Authorization number from the Customer Service department and then send it with ship ping costs prepaid When shipping any instrument make sure it is properly packaged for complete protection Han...

Страница 36: ...e using these products make sure that they are entirely suitable for the environment in which they are used Operation of these instruments in residential areas could cause unaccept able interferences...

Страница 37: ...4 2942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 South Africa Tel 011 615 6076...

Отзывы: