23
SETUP
Parameter
(Scrolled Message)
Range / Option
Default Settings
Language
En (English), ES (Spanish),
Fr (French), Pt (Portuguese),
nL (Dutch), dE (German)
En (English)
*Available range changes based upon other settings.
High alarm must be set higher than low alarm. If low alarm was set 5.00 mS/cm (2500 ppm for
0.5 TDS factor) then high alarm range is 5.10 to 10.00 mS/cm (2510 to 9900 ppm, respectively).
Control
Option:
Auto to enable or Off (oFF) to disable the control
Press one of the arrow keys for the controller settings to change from Auto to oFF and vice versa.
To run a ten-seconds pump test, long press the arrow keys together until the pump starts to run.
“CONTROL” message is scrolled on the bottom of the LCD screen.
Control Type
Option:
On Off (on/oFF) or Proportional (ProP)
Press the arrow keys to switch between on/off and proportional options.
“CONTROL
TYPE” message is scrolled on the bottom of the LCD screen.
Note:
To enter Control Type screen, pump control mode must be set as Auto.
Control Mode
Option:
Low (Lo) mode or High (Hi) mode
Press the arrow keys to switch between the options.
“CONTROL
MODE” message is scrolled on the bottom of the LCD screen.
Note:
To enter Control Mode screen, pump control mode must be set as Auto.
For more information see AUTOMATIC CONTROL TYPES in PUMP CONTROL section.